Intended use of the Xevo TQ-S cronos
Waters designed the Xevo TQ-S cronos for use as a research tool to accurately, reproducibly,
and robustly quantify target compounds present at the lowest possible levels in highly complex
sample matrices. The Xevo TQ-S cronos is not intended for use in diagnostic applications.
Calibrating
To calibrate LC systems, adopt acceptable calibration methods using at least five standards to
generate a standard curve. The concentration range for standards must include the entire range
of QC samples, typical specimens, and atypical specimens.
When calibrating mass spectrometers, consult the instrument’s online Help system for calibration
instructions.