4. For ESI probe adapters, connect the ESI probe adapter’s cable to the high-voltage
connector.
5. Install the probe assembly. See
Installing the probe assembly (Page 54)
.
3.1.6 Installing the probe assembly
Requirements:
• Ensure that you use a probe assembly appropriate for your application. For example, using
an ESI probe assembly with an APCI probe adapter compromises instrument performance.
Use the part number on the identification label to verify the probe assembly type.
• Ensure that the probe adapter is installed on the source, with the probe adapter cap removed.
Installing the probe adapter (Page 51)
.
• To minimize delays and dispersion, select the shortest probe assembly to connect the diverter
valve to the probe.
Notice:
Do not adjust the length of the probe assembly. Cutting the PEEKsil tubing
renders the probe assembly unusable.
Figure 3–10: Probe assembly
Capillary
Probe adapter PEEK fitting
Warning label
Identification label showing part number
Fitting spring retainer
Probe inlet fitting
Notes: