Figure 4–45: Fitting the seal into the groove
Seal
Groove
4. Fit the following new seals to the probe adjuster assembly:
• Probe seal
• Nebulizer gas seal
5. Fit the source enclosure to the instrument (see
Fitting the source enclosure to the
4.22 Replacing the air filter
You must replace the air filter annually.
Required materials
• New air filter
To replace the air filter:
1. Lift the instrument's visor so the source probe is fully exposed.
2. Fully open the source enclosure.
3. Disconnect the probe cable from the high-voltage connector, and leave the cable in a
position that does not obstruct the air filter grill.
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)