3 Maintenance Procedures
88
September 26, 2013, 715004208 Rev. A
2. Using the T10 TORX driver, loosen the two captive screws on the
restrictor module you are replacing, and then remove the module.
3. Insert the new restrictor module into the flow-splitter assembly.
4. Using the T10 TORX driver, tighten the captive screws on the restrictor
module to the extent possible.
5. Connect the inlet 1, outlet 1, inlet 2, and outlet 2 tubing to the new
restrictor module.
Tip:
To prevent contamination, install plugs on any unused restrictor
modules.
Replacing the restrictor module in the optional ACQUITY
UPC
2
Flow-splitter kit
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Restrictor module
Required tool
T10 TORX driver (startup kit)
To replace the restrictor module:
1. Disconnect inlet 3, outlet 2, and outlet 3 from the restrictor module.
Restrictor module (3)
Captive screws