3 Maintenance Procedures
126
September 26, 2013, 715004208 Rev. A
10. Using the 5/16-inch open-end wrench to hold the check-valve cartridge
in place, disconnect the tubing connection from the check valve with the
1/4-inch open-end wrench.
11. Disconnect the pressure transducer cable from the bulkhead by
squeezing on the tabs and pulling gently.
Accumulator check valve
Compression fitting
Place 5/16-inch open-end
wrench here
Pressure transducer
cable connector
Tabs