Removing the primary head and replacing its seals
September 26, 2013, 715004208 Rev. A
111
13. Using the T27 TORX driver, loosen and remove the 2 support plate
bolts, and then gently pull the head and support plate off the actuator
housing, making sure not to tilt the head during the extraction.
Pulling head and support plate off actuator housing:
Caution:
To avoid damaging the plunger, support the head from
below as you remove it.
Support plate bolt (2)
Head and support plate
Plunger