D e t e c t i o n

NEO Photo-ionization detector 

User Manual

D e t e c t i o n

26

10. Specifications

Size

230 x 74 x 55 mm   (9.1 x 2.9 x 2.2 in)  (with boot) 

Weight

708 g (24.9 oz)  (w/boot) 

Sensor

Photo-ionization sensor with standard 10.6 eV lamp (9.8 eV lamp in MP186)* 

Response time T

90

3 seconds

Accuracy 

±3% (at calibration point) 

Battery / Run time

Rechargeable Lithium-Ion battery with 24 hours typical operation 

Keypad

4 Operation keys 

Sampling Pump

Built-in pump with 3 settings from 300 to 430 cc/min Sample from up to 100 ft (30 m) 

Display

128 x 128 graphical LCD, 1.77 x 1.73 in (45 x 44 mm), with LED backlight for enhanced 
display readability 

Direct Readout 

Real-time reading of gas concentration (ppb, ppm, mg/m3, μg/m3), PID measurement 
gas and correction factor, lamp on/off, Man-Down alarm on/ off, battery  status, pump 
status, datalogging on/off, wireless 
on/off, temperature and time 

Alarms

Audible (95 dB @ 30 cm), visual (flashing bright red LEDs), and on-screen indication of 
alarm conditions plus wireless remote alarm notification 
High, Low, TWA and STEL alarms 
Over range alarm, battery low alarm  
Man-Down alarm with pre-alarm and real-time re-mote wireless notification 

Datalogging 

Capacity

Standard 12 months at one-minute intervals  Storage interval adjustable from 1 to 3,600 
seconds 

Calibration

Two/three-point calibration

Low Flow Alarm

Auto pump shutoff at low-flow condition

Charging and

communication

Charging, data download, instrument setup and firm-ware upgrades on PC or laptop via 
Micro USB Wireless data and alarm status transmission via built-in RF modem

Wireless range

1,000 ft (300 m) line of sight

Correction Factors

Integrated Correction Factor list of more than 200 compounds

IP Rating

IP-66 when turned on / IP-67 when turned off

EMI/RFI

Highly resistant to EMI/RFI 
Compliant with EMC Directive 2014/30/EU

Safety certifications

UL/cUL:

 Class I, Div 1, Group ABCD 

IECEx

: Ex ia IIC T4 Ga 

ATEX: 

II 1G  Ex ia IIC T4  Ga 

CE 

Conformité Européenne

Temperature

-20° to 50°C (-4° to 122°F) 

Humidity

0% to 95% Relative humidity (non-condensing)

Attachments

Durable rubber boot, color coded for different models; Tube holder for MP186

Warranty

2 Years including lamp and sensor (1-year for 9.8 eV lamp)

Содержание NEO MP181

Страница 1: ...www watchgas eu NEO Photo ionization Detector User Manual D e t e c t i o n...

Страница 2: ...m Signals 8 5 6 1 Alarm Signal Summary 8 5 6 2 Preset Alarm Limits Calibration 8 5 6 3 Testing the alarms 9 5 7 Backlight 9 5 8 Datalogging 9 5 9 Man Down 9 5 10 Wireless 9 6 Configuration Mode 10 6 1...

Страница 3: ...anufacturer s instructions The user should understand how to set the correct parameters and interpret the obtained results Use only WatchGas mPower battery pack M011 3002 000 Substitution of component...

Страница 4: ...strument s user interface consists of the LCD display Alarm LEDs an alarm transducer and four keys The keys are Esc Ent OK Esc Ent OK Esc Ent OK Esc Ent OK Left key Right Key Up Key Down Mode Key INLE...

Страница 5: ...WatchGas mPower USB cable P N M 011 3003 000 is required for a PC to recognize the instrument and communicate with mPower Suite Item Description Gas information Shows the measurement gas and its corre...

Страница 6: ...nt is preset with default alarm limits and the sensor is pre calibrated with standard calibration gas However the instrument should be tested and the calibration verified before the first use After th...

Страница 7: ...the startup procedure is complete and a numerical reading screen with icons is displayed the instrument is fully functional and ready for use FLOW CHART FOR BASIC USER MODE 5 4 INTEGRATED SAMPLING PU...

Страница 8: ...the preset limits the loud buzzer and red flashing LED are activated immediately to warn of the alarm condition In addition the instrument alarms if one of the following conditions occurs battery vol...

Страница 9: ...site ID serial number last calibration date and alarm limits All data are retained even after the unit is turned off in non volatile memory so that it can be down loaded at a later time to a PC Stored...

Страница 10: ...pressing the Up key Esc Ent OK or Mode key Esc Ent OK Press the Right key Esc Ent OK or the Left key Esc Ent OK to move the cursor After entering the desired numbers press the Right key Esc Ent OK or...

Страница 11: ...hlighted Use the Up key Esc Ent OK or Mode key Esc Ent OK to scroll through the menu until the desired item is reached and use the Right key to select it To Exit Config Mode and return to normal opera...

Страница 12: ...D e t e c t i o n NEO Photo ionization detector User Manual D e t e c t i o n 12 FLOW CHART FOR FOR CONFIGURATION MODE...

Страница 13: ...trument pump Alternatively the span gas can first be filled into a gas bag e g Tedlar bag or delivered through a demand flow regulator to match the pump flow precisely Another alternative is to use a...

Страница 14: ...pressing the Right key to enter the submenu The message Please Apply Gas is displayed Start the span calibration by pressing the Up key which begins a 30 second countdown Note To abort the Span calibr...

Страница 15: ...es with the instrument setting 6 3 2 MEASUREMENT The sub menus for Measurement are Measurement Unit and Measurement Gas Measurement Unit Standard available measurement units include Measurement Gas Me...

Страница 16: ...eft key to move the cursor Move the cursor to after the setting is complete and press Up to save and exit Low STEL and TWA Alarms Set the Low STEL and TWA Alarm limits in the same manner as described...

Страница 17: ...ght key to enter Man down alarm menu Man Down Alarm Off On Press Right to make a selection followed by Left to save and exit Motionless Time Motionless Time is the trigger time for Man Down alarm when...

Страница 18: ...CAUTION Once the data are cleared they cannot be recovered 1 Press the Right key to clear the datalog The display asks Clear Datalog 2 Press the Left key if you do not want to clear the datalog 3 Pre...

Страница 19: ...n move the cursor to and press Up to save and exit Backlight The Backlight can be set as Automatic Manual or Off In Automatic Mode the backlight comes on and stays on in low light conditions In Manual...

Страница 20: ...t calibration Zero Span 1 and Span 2 can be enabled for more accuracy particularly in the high concentration range above about 500 ppm where response is less linear Use the Right key to enter 3 Point...

Страница 21: ...the lamp ages 2 a dust or oil film accumulates on the lamp window or 3 a change in environmental conditions such as background matrix gas humidity or temperature The default setting of self zeroing is...

Страница 22: ...Connected list Click on the S N to get the configuration file from the instrument 5 Edit the configuration parameters as desired and click Write to upload the configuration to the instrument 6 Read a...

Страница 23: ...he screen Below is a sample screen of datalog information listing sample point time and instantaneous reading A new Single Datalog file is created each time the instrument is turned on or the configur...

Страница 24: ...bed below 3 Use a cotton swab wetted with methanol to clean the flat window surface of the lamp If greasy dirt is hard to remove using methanol the window can be polished using fine alumina powder pol...

Страница 25: ...Dirty Filter Replace filter Dirty sensor module Clean or replace sensor module Excessive moisture or water condensation Blow dry the sensor module Incorrect calibration Calibrate the unit Reading abn...

Страница 26: ...s remote alarm notification High Low TWA and STEL alarms Over range alarm battery low alarm Man Down alarm with pre alarm and real time re mote wireless notification Datalogging Capacity Standard 12 m...

Страница 27: ...ing damage by fire lightening water damage or other hazard voids liability of the manufacturer In the event that a product should fail to perform up to manufacturer specifications during the appli cab...

Отзывы: