3.Charging a Lithium-Ion Battery 

Always fully charge the battery before using the instrument. The screen will display a battery icon from 

empty (no bars) to fully charged (3 bars). Follow this procedure to charge the instrument: 

1.  Plug the Micro-USB into the charging port at the bottom of the instrument.

2.  Plug the USB connector* into either a) a personal computer, or b) into the AC/DC adapter and plug 

the AC/DC adapter into the wall outlet.

The instrument begins charging automatically. The charging LED turns red to indicate  charging. 

During charging, the 3 bars in the battery icon on the instrument’s display are animated. When 

the battery is fully charged, the LED turns green, the icon is no longer animated and shows a full 

battery, and the message ‘Fully Charged’ is displayed.

*NOTE: Any locally-obtained USB A to Micro B USB cable will work for charging, but will not work for 

communication with mPower Suite configuration and data transfer software. The WatchGas mPower 

USB cable P/N M-011-3003-000

 

is required for a PC to recognize the instrument and communicate 

with mPower Suite.

Item

Description

Gas information 

Shows the measurement gas and its correction factor 

Reading

Concentration of gas measured by the instrument 

Calibration/Bump Due

Indicates if calibration or bump status is OK or not 

Pump Status

Indicates whether the pump is working or blocked 

Man Down Status

Indicates if Man Down alarm is on

Lamp Status

Indicates if the lamp is on or not

Datalog Status

Indicates whether datalog is on or off

Battery Status

Indicates battery capacity in 3 bars

Time 

Indicates current time

Temperature

Indicates current temperature

Warning

To reduce the risk of ignition of hazardous atmospheres, recharge and replace batteries only 
in areas known to be non-hazardous. Use only  WatchGas mPower’s rechargeable lithium 
battery part number: M011-3002-000.

3.1 LOW VOLTAGE WARNING

When the battery’s charge falls below a preset voltage, the instrument warns you by beeping 

once and flashing once every minute, and the “empty battery” icon blinks on and off once per 

second. Turn off the instrument within 10 minutes and either recharge the battery, or replace the 

battery with a fresh one with a full charge.

Содержание NEO MP181

Страница 1: ...www watchgas eu NEO Photo ionization Detector User Manual D e t e c t i o n...

Страница 2: ...m Signals 8 5 6 1 Alarm Signal Summary 8 5 6 2 Preset Alarm Limits Calibration 8 5 6 3 Testing the alarms 9 5 7 Backlight 9 5 8 Datalogging 9 5 9 Man Down 9 5 10 Wireless 9 6 Configuration Mode 10 6 1...

Страница 3: ...anufacturer s instructions The user should understand how to set the correct parameters and interpret the obtained results Use only WatchGas mPower battery pack M011 3002 000 Substitution of component...

Страница 4: ...strument s user interface consists of the LCD display Alarm LEDs an alarm transducer and four keys The keys are Esc Ent OK Esc Ent OK Esc Ent OK Esc Ent OK Left key Right Key Up Key Down Mode Key INLE...

Страница 5: ...WatchGas mPower USB cable P N M 011 3003 000 is required for a PC to recognize the instrument and communicate with mPower Suite Item Description Gas information Shows the measurement gas and its corre...

Страница 6: ...nt is preset with default alarm limits and the sensor is pre calibrated with standard calibration gas However the instrument should be tested and the calibration verified before the first use After th...

Страница 7: ...the startup procedure is complete and a numerical reading screen with icons is displayed the instrument is fully functional and ready for use FLOW CHART FOR BASIC USER MODE 5 4 INTEGRATED SAMPLING PU...

Страница 8: ...the preset limits the loud buzzer and red flashing LED are activated immediately to warn of the alarm condition In addition the instrument alarms if one of the following conditions occurs battery vol...

Страница 9: ...site ID serial number last calibration date and alarm limits All data are retained even after the unit is turned off in non volatile memory so that it can be down loaded at a later time to a PC Stored...

Страница 10: ...pressing the Up key Esc Ent OK or Mode key Esc Ent OK Press the Right key Esc Ent OK or the Left key Esc Ent OK to move the cursor After entering the desired numbers press the Right key Esc Ent OK or...

Страница 11: ...hlighted Use the Up key Esc Ent OK or Mode key Esc Ent OK to scroll through the menu until the desired item is reached and use the Right key to select it To Exit Config Mode and return to normal opera...

Страница 12: ...D e t e c t i o n NEO Photo ionization detector User Manual D e t e c t i o n 12 FLOW CHART FOR FOR CONFIGURATION MODE...

Страница 13: ...trument pump Alternatively the span gas can first be filled into a gas bag e g Tedlar bag or delivered through a demand flow regulator to match the pump flow precisely Another alternative is to use a...

Страница 14: ...pressing the Right key to enter the submenu The message Please Apply Gas is displayed Start the span calibration by pressing the Up key which begins a 30 second countdown Note To abort the Span calibr...

Страница 15: ...es with the instrument setting 6 3 2 MEASUREMENT The sub menus for Measurement are Measurement Unit and Measurement Gas Measurement Unit Standard available measurement units include Measurement Gas Me...

Страница 16: ...eft key to move the cursor Move the cursor to after the setting is complete and press Up to save and exit Low STEL and TWA Alarms Set the Low STEL and TWA Alarm limits in the same manner as described...

Страница 17: ...ght key to enter Man down alarm menu Man Down Alarm Off On Press Right to make a selection followed by Left to save and exit Motionless Time Motionless Time is the trigger time for Man Down alarm when...

Страница 18: ...CAUTION Once the data are cleared they cannot be recovered 1 Press the Right key to clear the datalog The display asks Clear Datalog 2 Press the Left key if you do not want to clear the datalog 3 Pre...

Страница 19: ...n move the cursor to and press Up to save and exit Backlight The Backlight can be set as Automatic Manual or Off In Automatic Mode the backlight comes on and stays on in low light conditions In Manual...

Страница 20: ...t calibration Zero Span 1 and Span 2 can be enabled for more accuracy particularly in the high concentration range above about 500 ppm where response is less linear Use the Right key to enter 3 Point...

Страница 21: ...the lamp ages 2 a dust or oil film accumulates on the lamp window or 3 a change in environmental conditions such as background matrix gas humidity or temperature The default setting of self zeroing is...

Страница 22: ...Connected list Click on the S N to get the configuration file from the instrument 5 Edit the configuration parameters as desired and click Write to upload the configuration to the instrument 6 Read a...

Страница 23: ...he screen Below is a sample screen of datalog information listing sample point time and instantaneous reading A new Single Datalog file is created each time the instrument is turned on or the configur...

Страница 24: ...bed below 3 Use a cotton swab wetted with methanol to clean the flat window surface of the lamp If greasy dirt is hard to remove using methanol the window can be polished using fine alumina powder pol...

Страница 25: ...Dirty Filter Replace filter Dirty sensor module Clean or replace sensor module Excessive moisture or water condensation Blow dry the sensor module Incorrect calibration Calibrate the unit Reading abn...

Страница 26: ...s remote alarm notification High Low TWA and STEL alarms Over range alarm battery low alarm Man Down alarm with pre alarm and real time re mote wireless notification Datalogging Capacity Standard 12 m...

Страница 27: ...ing damage by fire lightening water damage or other hazard voids liability of the manufacturer In the event that a product should fail to perform up to manufacturer specifications during the appli cab...

Отзывы: