background image

www.watchgas.eu

D e t e c t i o n

User Manual 

Photo-ionization detector NEO

D e t e c t i o n

13

6.3.1 CALIBRATION

Calibration Overview

Although the instrument will typically hold an approximate calibration for 

several weeks, when used for worker health and safety it should be calibrated or 

bump tested before each day’s use. The interval between span calibrations can 

be extended based on the user’s experience for how long the calibration holds 

under their operating conditions. For best accuracy the calibration gas chosen 

is the same as the gas to be measured. However, this is often inconvenient, and 

isobutylene has become the standard surrogate calibration gas because it has 

good response and is inexpensive, stable, and non-toxic. A correction factor (CF) 

can then be applied to make the NEO display in true concentration units of the 

gas being measured. This CF is selected from the instrument’s gas library of over 

200 compounds, or entered manually.

Calibration Set-Up

Span gas from a cylinder is conveniently supplied to the instrument using a regulator with fixed flow 

rate of about 0.5L/min (500 cc per min.) to match or slightly exceed the flow rate of the instrument 

pump. Alternatively, the span gas can first be filled into a gas bag (e.g., Tedlar® bag) or delivered through 

a demand-flow regulator to match the pump flow precisely. Another alternative is to use a regulator 

with >500 cc/min flow but allow the excess flow to escape through a T connector or an open tube.

Zero Calibration

This procedure determines the zero point of the sensor calibration curve. Connect the instrument to a 

clean air source free of VOCs such as from a cylinder, gas bag, or ambient air filtered through a charcoal 

tube (VOC Zeroing Tube). Most outdoor air is sufficiently free of contaminants that it can be used for 

zero calibration except for measurements in the low ppb range. The air source should have an oxygen 

concentration of 20.9% (or the same as in the gas to be measured) because oxygen levels have some 

effect on the PID response.

Select ‘Zero Calib’ by pressing the Right key to enter the submenu. The message ‘ Please Apply Zero Gas’ 

is displayed. Start the zero calibration by pressing the Up key, and a 30- second countdown begins.

Note: To abort the zeroing and proceed to Span calibration, press the Left key at any time during the 

process. A confirmation message ‘Zero aborted!’ will appear, followed by the Span calibration menu.

When complete, this message “Zero Calibration Done!” is shown, and the display automatically moves to 

the Span Calibration menu.

Содержание NEO MP181

Страница 1: ...www watchgas eu NEO Photo ionization Detector User Manual D e t e c t i o n...

Страница 2: ...m Signals 8 5 6 1 Alarm Signal Summary 8 5 6 2 Preset Alarm Limits Calibration 8 5 6 3 Testing the alarms 9 5 7 Backlight 9 5 8 Datalogging 9 5 9 Man Down 9 5 10 Wireless 9 6 Configuration Mode 10 6 1...

Страница 3: ...anufacturer s instructions The user should understand how to set the correct parameters and interpret the obtained results Use only WatchGas mPower battery pack M011 3002 000 Substitution of component...

Страница 4: ...strument s user interface consists of the LCD display Alarm LEDs an alarm transducer and four keys The keys are Esc Ent OK Esc Ent OK Esc Ent OK Esc Ent OK Left key Right Key Up Key Down Mode Key INLE...

Страница 5: ...WatchGas mPower USB cable P N M 011 3003 000 is required for a PC to recognize the instrument and communicate with mPower Suite Item Description Gas information Shows the measurement gas and its corre...

Страница 6: ...nt is preset with default alarm limits and the sensor is pre calibrated with standard calibration gas However the instrument should be tested and the calibration verified before the first use After th...

Страница 7: ...the startup procedure is complete and a numerical reading screen with icons is displayed the instrument is fully functional and ready for use FLOW CHART FOR BASIC USER MODE 5 4 INTEGRATED SAMPLING PU...

Страница 8: ...the preset limits the loud buzzer and red flashing LED are activated immediately to warn of the alarm condition In addition the instrument alarms if one of the following conditions occurs battery vol...

Страница 9: ...site ID serial number last calibration date and alarm limits All data are retained even after the unit is turned off in non volatile memory so that it can be down loaded at a later time to a PC Stored...

Страница 10: ...pressing the Up key Esc Ent OK or Mode key Esc Ent OK Press the Right key Esc Ent OK or the Left key Esc Ent OK to move the cursor After entering the desired numbers press the Right key Esc Ent OK or...

Страница 11: ...hlighted Use the Up key Esc Ent OK or Mode key Esc Ent OK to scroll through the menu until the desired item is reached and use the Right key to select it To Exit Config Mode and return to normal opera...

Страница 12: ...D e t e c t i o n NEO Photo ionization detector User Manual D e t e c t i o n 12 FLOW CHART FOR FOR CONFIGURATION MODE...

Страница 13: ...trument pump Alternatively the span gas can first be filled into a gas bag e g Tedlar bag or delivered through a demand flow regulator to match the pump flow precisely Another alternative is to use a...

Страница 14: ...pressing the Right key to enter the submenu The message Please Apply Gas is displayed Start the span calibration by pressing the Up key which begins a 30 second countdown Note To abort the Span calibr...

Страница 15: ...es with the instrument setting 6 3 2 MEASUREMENT The sub menus for Measurement are Measurement Unit and Measurement Gas Measurement Unit Standard available measurement units include Measurement Gas Me...

Страница 16: ...eft key to move the cursor Move the cursor to after the setting is complete and press Up to save and exit Low STEL and TWA Alarms Set the Low STEL and TWA Alarm limits in the same manner as described...

Страница 17: ...ght key to enter Man down alarm menu Man Down Alarm Off On Press Right to make a selection followed by Left to save and exit Motionless Time Motionless Time is the trigger time for Man Down alarm when...

Страница 18: ...CAUTION Once the data are cleared they cannot be recovered 1 Press the Right key to clear the datalog The display asks Clear Datalog 2 Press the Left key if you do not want to clear the datalog 3 Pre...

Страница 19: ...n move the cursor to and press Up to save and exit Backlight The Backlight can be set as Automatic Manual or Off In Automatic Mode the backlight comes on and stays on in low light conditions In Manual...

Страница 20: ...t calibration Zero Span 1 and Span 2 can be enabled for more accuracy particularly in the high concentration range above about 500 ppm where response is less linear Use the Right key to enter 3 Point...

Страница 21: ...the lamp ages 2 a dust or oil film accumulates on the lamp window or 3 a change in environmental conditions such as background matrix gas humidity or temperature The default setting of self zeroing is...

Страница 22: ...Connected list Click on the S N to get the configuration file from the instrument 5 Edit the configuration parameters as desired and click Write to upload the configuration to the instrument 6 Read a...

Страница 23: ...he screen Below is a sample screen of datalog information listing sample point time and instantaneous reading A new Single Datalog file is created each time the instrument is turned on or the configur...

Страница 24: ...bed below 3 Use a cotton swab wetted with methanol to clean the flat window surface of the lamp If greasy dirt is hard to remove using methanol the window can be polished using fine alumina powder pol...

Страница 25: ...Dirty Filter Replace filter Dirty sensor module Clean or replace sensor module Excessive moisture or water condensation Blow dry the sensor module Incorrect calibration Calibrate the unit Reading abn...

Страница 26: ...s remote alarm notification High Low TWA and STEL alarms Over range alarm battery low alarm Man Down alarm with pre alarm and real time re mote wireless notification Datalogging Capacity Standard 12 m...

Страница 27: ...ing damage by fire lightening water damage or other hazard voids liability of the manufacturer In the event that a product should fail to perform up to manufacturer specifications during the appli cab...

Отзывы: