3.2 Equipment location
The equipment is designed for indoor use and < 3000 m altitude.
The location of the equipment must not obstruct the access to the electric connection.
Make sure that the rear connections of the equipment are not adversely affected by an inappropriate location.
An easily accessible stopcock must be installed between the tapping point and the equipment inlet in case the water supply
must be cut.
The equipment needs a drainage to collect the rejection water.
3.3 Equipment features:
Production flow *:
2.5 l/h *
Operating pressure:
2.5 - 6 bar
Conductivity
(compensated at 25°C)
< 0.2
µ
S/cm
Silica removal:
>99.9 %
Reverse osmosis performance: 90 - 98%
Weight in operation
12 Kg
* 15% fluctuation. The production values are based on standard operating conditions [25 °C / pH 5 - 7 / 4 bar/250 ppm NaCl] and
may vary depending on the feed water quality and typical variations in the reverse osmosis membranes.
Equipment dimensions:
(*) Reservoir is optional