Connect the Micromatic with the Level connector using a wire supplied with the equipment.
d. Water Type II: Connect a tube at the rear fitting 2 with the fitting placed on the cap of the tank.
7.3 Equipment start-up
Once the hydraulic connections have been made, open the stopcock and check for leaks at the connections made, including the
dispensing faucet.
Electrical connection:
First connect the transformer plug-in to the equipment.
Then
connect the transformer to the power system: The equipment will start automatically.
Leave the equipment running until the tank is full.
The screen displays the conductivity value of the water produced.
When the tank is full, the equipment stops the production of water and displays the full tank icon.
The screen displays the conductivity value of the water produced at the time the equipment stopped.