2. Safety Information  

The security information included in this chapter should be read carefully.   
  
The user should check for compliance with security conditions set out in this Manual to ensure the safe use of the equipment.   

  

In case of doubt, contact the Technical Assistance Service.  
  
Safety information is related to both user protection and equipment protection.   
  

   Access to the equipment interior 

  
Only  authorised  technical  service  may  access  the  equipment  interior  to  carry  out  maintenance  and  repair  tasks  other  than  the 
routine replacement of expendable items.  

  

  Electrical connections  

  

-  Check that connection to the power system is 110 - 230 VAC, 50 -60 Hz.   
-  Only use the transformer supplied with the equipment. Plug Type C, Europlug. 
-  Supply voltage fluctuations should not exceed 10% of the rated voltage.  
-  Supported  transitional  overvoltage:  The  typical  present  in  the  power  supply  network  (category  II  of  the  standard  IEC 

60634-4-443).  

-  The electrical plug should not be situated below the equipment.   
-  The electrical plug should be accessible in case of be necessary to disconnect the equipment quickly.  

 

   Explosive atmospheres  

The equipment is not designed to work in explosive atmospheres.  
  

  

 Environmental pollution permitted  

Pollution grade 1 (No environmental pollution or only dry pollution, not conductive).  

  

  Feed water and hydraulic connections  

  
The feed water must comply with the specifications listed in this Manual.  
Feeding  poor  quality  water  (turbidity  and/or  high  hardness,  inadequate  conductivity,  unwanted  pollutants,  etc.)  can  result  in 
deterioration of the equipment components or a lower performance.  
  
Respect the pressure conditions established for the feed water to the equipment.  High pressure can damage the hydraulic system.  
In these cases, it is necessary to install a pressure regulator before the equipment.  
  
  
 A feed  water  pressure  lower  than  the minimum  required  will  cause  poorer  performance  and  improper  functioning  of some  of the 
system components.  
  
Have a stopcock that is easily accessible when you need to cut the water supply to the equipment.  
  

      Locate  the  equipment  so  that  the  electric  and  hydraulic  connections  located  on  the  rear  of  the  equipment  are  not 

adversely affected.   
Prevent straining the power cable or hydraulic connections.  
  
 

  

  Consumables  

  
Replace the consumables items in the times anticipated and based on the messages from your equipment.   
Use only original expendable items.  Using non original components may cause equipment damage and loss of guarantee.  

  

 To move the equipment, handle it by holding at the base.  

  

Содержание Micromatic

Страница 1: ...Micromatic DA Equipment User s Manual Rev 7 20 03 17...

Страница 2: ...as well as its electrical and hydraulic connection aspects The User Manual also contains additional information about the equipment and settings allowing a better understanding of the operation and ca...

Страница 3: ...description 9 5 1 Pretreatment 9 5 2 Reverse osmosis 9 5 3 Deionisation 9 5 4 Type II Water Storage 9 5 5 Automatic operation of the equipment 9 6 Hydraulic diagram 10 7 Installation of the equipment...

Страница 4: ...ive atmospheres Environmental pollution permitted Pollution grade 1 No environmental pollution or only dry pollution not conductive Feed water and hydraulic connections The feed water must comply with...

Страница 5: ...wer point should not be situated below the equipment Power consumption 24 VA System voltage 24 Vdc Feed water requirements Drinking water according to European or USA standards Maximum Inlet pressure...

Страница 6: ...ment inlet in case the water supply must be cut The equipment needs a drainage to collect the rejection water 3 3 Equipment features Production flow 2 5 l h Operating pressure 2 5 6 bar Conductivity c...

Страница 7: ...Connection with open tank level detector 1 Rejection Rejection water outlet from the reverse osmosis phase directed to a drainage 2 Tank Water outlet from the equipment to a Type II water storage tank...

Страница 8: ...commands Scrolling the menu Value decrease Scrolling the menu Value increase 4 5 3 Acoustic advice The equipment beeps when activating any warning Pressing any key turns off the sound 4 6 Type II Wate...

Страница 9: ...with high inlet pressure 6 bar it is necessary to install a pressure regulator before the water inlet to the equipment 5 3 Deionisation The permeate passes through a mixed bed of strong cation anion r...

Страница 10: ...ner wrench Pipe cutter 1 set of pretreatment cartridges 1 and 2 1 set of deionization resin cartridges 3 4 Tank with level optional or Level User Manual Water inlet and drainage connection kit The equ...

Страница 11: ...et Electrical connection Transformer Power connection 110 230 VAC Socket Type C Europlug Connection to the equipment Rear connection Plug in type Accessible plug in case of need to disconnect the powe...

Страница 12: ...ed Do not use Teflon for this connection the connection is fitted with a rubber seal If due to highly polluted inlet water it is necessary to install and additional pretreatment the connection sequenc...

Страница 13: ...made including the dispensing faucet Electrical connection First connect the transformer plug in to the equipment Then connect the transformer to the power system The equipment will start automaticall...

Страница 14: ...d sent to the tank The icons Equipment in Operation Electrovalve Tank full are activated 8 1 Symbols and meaning Pre treatment change Equipment in operation Deionisation Resin change Tank full Water c...

Страница 15: ...pressing the keys and Parameter Value uS Value 10 0 S cm Hour setting 180 Equipment operation hours and time RESET XXX Temperature value 25 0 C To modify the set values take the following steps a Sel...

Страница 16: ...25 C 8 3 3 Resetting the pretreatment hour meter When the pretreatment is changed it is necessary to reset the hour meter 1 Access the set up menu Press MODE ESC 2 Press 3 The screen displays the equi...

Страница 17: ...3 Turn on the water supply to the equipment 9 2 Resin cartridge replacement Follow the same procedures as when replacing the pretreatment cartridges 9 3 RO membrane replacement The RO membrane should...

Страница 18: ...ivered adjusted and calibrated ex works The equipment can be calibrated and adjusted at any time by the Technical Assistance following a calibration pattern according to international standards It is...

Страница 19: ...re The lower it is the lower the performance of the equipment Release the pretreatment cartridges and check the degree of fouling If they are very dirty the silting blocks the flow of water Check that...

Страница 20: ...0 1 2002 and 61010 1 Corr 2003 Security requirements for electronic measuring control and laboratory use equipments 2004 108 CE Electromagnetic Compatibility Directive Royal Decree 1580 2006 UNE EN 61...

Отзывы: