Micromatic DA 

Equipment

  

User´s Manual 

 

  

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Rev. 7     

    20.03.17

  

  

Содержание Micromatic

Страница 1: ...Micromatic DA Equipment User s Manual Rev 7 20 03 17...

Страница 2: ...as well as its electrical and hydraulic connection aspects The User Manual also contains additional information about the equipment and settings allowing a better understanding of the operation and ca...

Страница 3: ...description 9 5 1 Pretreatment 9 5 2 Reverse osmosis 9 5 3 Deionisation 9 5 4 Type II Water Storage 9 5 5 Automatic operation of the equipment 9 6 Hydraulic diagram 10 7 Installation of the equipment...

Страница 4: ...ive atmospheres Environmental pollution permitted Pollution grade 1 No environmental pollution or only dry pollution not conductive Feed water and hydraulic connections The feed water must comply with...

Страница 5: ...wer point should not be situated below the equipment Power consumption 24 VA System voltage 24 Vdc Feed water requirements Drinking water according to European or USA standards Maximum Inlet pressure...

Страница 6: ...ment inlet in case the water supply must be cut The equipment needs a drainage to collect the rejection water 3 3 Equipment features Production flow 2 5 l h Operating pressure 2 5 6 bar Conductivity c...

Страница 7: ...Connection with open tank level detector 1 Rejection Rejection water outlet from the reverse osmosis phase directed to a drainage 2 Tank Water outlet from the equipment to a Type II water storage tank...

Страница 8: ...commands Scrolling the menu Value decrease Scrolling the menu Value increase 4 5 3 Acoustic advice The equipment beeps when activating any warning Pressing any key turns off the sound 4 6 Type II Wate...

Страница 9: ...with high inlet pressure 6 bar it is necessary to install a pressure regulator before the water inlet to the equipment 5 3 Deionisation The permeate passes through a mixed bed of strong cation anion r...

Страница 10: ...ner wrench Pipe cutter 1 set of pretreatment cartridges 1 and 2 1 set of deionization resin cartridges 3 4 Tank with level optional or Level User Manual Water inlet and drainage connection kit The equ...

Страница 11: ...et Electrical connection Transformer Power connection 110 230 VAC Socket Type C Europlug Connection to the equipment Rear connection Plug in type Accessible plug in case of need to disconnect the powe...

Страница 12: ...ed Do not use Teflon for this connection the connection is fitted with a rubber seal If due to highly polluted inlet water it is necessary to install and additional pretreatment the connection sequenc...

Страница 13: ...made including the dispensing faucet Electrical connection First connect the transformer plug in to the equipment Then connect the transformer to the power system The equipment will start automaticall...

Страница 14: ...d sent to the tank The icons Equipment in Operation Electrovalve Tank full are activated 8 1 Symbols and meaning Pre treatment change Equipment in operation Deionisation Resin change Tank full Water c...

Страница 15: ...pressing the keys and Parameter Value uS Value 10 0 S cm Hour setting 180 Equipment operation hours and time RESET XXX Temperature value 25 0 C To modify the set values take the following steps a Sel...

Страница 16: ...25 C 8 3 3 Resetting the pretreatment hour meter When the pretreatment is changed it is necessary to reset the hour meter 1 Access the set up menu Press MODE ESC 2 Press 3 The screen displays the equi...

Страница 17: ...3 Turn on the water supply to the equipment 9 2 Resin cartridge replacement Follow the same procedures as when replacing the pretreatment cartridges 9 3 RO membrane replacement The RO membrane should...

Страница 18: ...ivered adjusted and calibrated ex works The equipment can be calibrated and adjusted at any time by the Technical Assistance following a calibration pattern according to international standards It is...

Страница 19: ...re The lower it is the lower the performance of the equipment Release the pretreatment cartridges and check the degree of fouling If they are very dirty the silting blocks the flow of water Check that...

Страница 20: ...0 1 2002 and 61010 1 Corr 2003 Security requirements for electronic measuring control and laboratory use equipments 2004 108 CE Electromagnetic Compatibility Directive Royal Decree 1580 2006 UNE EN 61...