SETUP  

The 

CL-200A

 has been designed primarily to drive Warner’s expanding line of Peltier-driven 

heating and cooling devices. However, the instrument is also capable of supplying power to Warner’s 
line of resistive heaters (Series 20 heater platforms, in-line solution heaters, etc.) by using the 

ACC-1

 

adapter cable at the peripheral device output (Item W). The instrument will auto-detect when a 
resistive-only heating device has been attached and will disable cooling commands.  

Warner Peltier Driven Systems  

The set up for the 

CL-200A

 

is straightforward when used with a Peltier driven heater/cooler. 

First connect the cable from the heater/cooler peripheral device into the 15-pin D-connector on the 
back of the 

CL-200A

. Then connect the 

TA-29

 thermistor (supplied with the heater/cooler) to the 

MONITOR TEMP IN

 BNC (Item F).  

If using the 

SC-20

 in-line solution heater/cooler, set the 

CL-200A 

loop speed to 

medium

. If using 

a different device, then set the loop speed to the setting specified in that devices user’s manual, or 
initially set the loop speed to 

fast

 and adjust as necessary to achieve stable performance. 

If using external control inputs, make connections from your 

command output

 (e.g., analog out 

on the A/D board or a power supply) to either the 

EXT TEMP SET

 (Item U) or 

EXT VOLT SET

 (Item T) 

BNC’s on the rear panel. 

Operating With Other Equipment  

The 

CL-200A

 can be used to power other Warner equipment such as the 

SH-27B

 and 

SF-28

 

inline solution heaters. In addition, the controller can be used as an independent device to power 
third-party equipment so long as the following considerations are met. 

Warner Resistive Heating Systems 

Use of the 

CL-200A

 to power Warner resistive heaters is permitted if used in conjunction 

with the 

ACC-1

 adapter cable. When the 

CL-200A

 is connected to a resistive device, all 

command inputs (auto and manual modes for both internal and external commands) are 
functional with the exception that cooling commands are not executed. 

Third-party Equipment 

The 

CL-200A

 heater controller will work with many other heating devices if used in 

conjunction with the 

ACC-1

 adapter cable. To obtain maximum heating power, the resistance of 

the third-party heating element should be between 8-12 

. However, any element that works at 

a maximum of 15 V and 2.4 A will also work with the 

CL-200A

.  

Thermistor Considerations  

The 

CL-200A

 heater controller is designed to accommodate Unical NTC thermistors from 

Thermometrics (Edison, New Jersey). This family of thermistors can be interchanged without 
recalibration of the instrument. The nominal resistance of a Unical NTC thermistor is 10.0 k

 

at 25

C. Other NTC thermistors may be used with the 

CL-200A

 if the nominal resistance is also 

10.0 k

 at 25

C.  

 

                         

                      

Warner

 Instruments 

Harvard Apparatus Company     

Содержание CL-200A

Страница 1: ...e en nt ts s D Du ua al l C Ch ha an nn ne el l B Bi ip po ol la ar r T Te em mp pe er ra at tu ur re e C Co on nt tr ro ol ll le er r M Mo od de el l C CL L 2 20 00 0A A Warner Instruments 1125 Dixwell Avenue Hamden CT 06514 800 599 4203 203 776 0664 203 776 1278 fax ...

Страница 2: ... Voltage Control Item K 7 Peltier Limit LED Item L 8 Freeze Alert LED Item M 8 Heat Cool LED s Item N 8 Current Limit LED Item O 8 Feedback Thermistor Selector Switch Item R 8 Thermistor Open LED Item S 8 Rear panel 9 External Volts Set Item T 9 External Temp Set Item U 9 Heater Cooler input Item W 9 Power Entry Module Item Y 9 Ground Terminals Item X 9 SETUP 10 Warner Peltier Driven Systems 10 Op...

Страница 3: ...CL 200A Manual Rev 120523 APPENDIX 13 Specifications 13 Accessories and replacement parts 13 Warranty and service 14 Warranty 14 Service notes 14 Certifications 15 ...

Страница 4: ...ple to use with a single control for temperature adjustment Total automatic control is provided in automatic mode while manual control is available in manual mode A loop speed selector is used to optimize the response of the system to accommodate thermal delay characteristics intrinsic to the setup Features of the CL 200A include Quiet operation suitable for use in sensitive electrophysiology appl...

Страница 5: ... goal is to provide clarity rather than complexity we welcome any feedback you may wish to provide Warner Instrument product model numbers are presented using bold type References to controls are specified using SMALL CAPS References to control settings are specified using italic type Special comments and warnings are presented in highlighted text Any other formatting should be apparent from conte...

Страница 6: ...w Set Temp reports the adjustment from either the SET TEMPERATURE control Item H or the EXT TEMP SET BNC Item Q Displayed units are C Control Temp reports the actual temperature of the feedback thermistor situated within the heater cooler device being used Displayed data is informative only and is in units of C Monitor Temp reports the temperature of the sensor thermistor which connects via the MO...

Страница 7: ...CL 200A is used in automatic mode Feedback options include fast medium and slow Optimally set to fast for most applications but can be set to medium or slow for systems with longer thermal delay characteristics For example use the medium speed for use with the SC 20 Bipolar In Line Solution Heater Set Temperature Control Item H This dial adjusts the set point of the automatic temperature control s...

Страница 8: ...E LED is lit when the Peltier device within the SC 20 approaches 0 C Activation of this circuit is informative only and will not alter any instrument setting Heat Cool LED s Item N RED and BLUE LED S indicate heating or cooling power applied to the Peltier device Current Limit LED Item O This indicator light informs the user that the channel has reached the maximum current output of 5 Amperes This...

Страница 9: ...A Input functions only when the INTERNAL EXTERNAL SELECT CONTROL Item J is set to external and the COOLING MODE SWITCH Item C is set to auto Units are 100 mV C This input can be used for computer control purposes Heater Cooler Peripheral Device Output Item W This is the connection port for Warner Heater Cooler devices such as the SC 20 Dual In line Heater Cooler and QE 1HC Heating Cooling Platform...

Страница 10: ...Item U or EXT VOLT SET Item T BNC s on the rear panel Operating With Other Equipment The CL 200A can be used to power other Warner equipment such as the SH 27B and SF 28 inline solution heaters In addition the controller can be used as an independent device to power third party equipment so long as the following considerations are met Warner Resistive Heating Systems Use of the CL 200A to power Wa...

Страница 11: ...th feedback thermistors can be used in automatic mode provided they are compatible with the CL 200A requirements LOOP SPEED is normally set to fast to provide the shortest cycle time between the application of power to the heater element and the sensing of temperature at the thermistor For heater cooler thermistor systems with long response times such as with Warner s SC 20 in line solution heater...

Страница 12: ...nto manual mode by switching the COOLING MODE SWITCH Item C to manual Switch the METER Item B to Output Volts and adjust to the desired set point using the MANUAL VOLTAGE control Item K Alternatively the CL 200A can be commanded using external inputs Set the CL 200A to external commands by setting the INTERNAL EXTERNAL SELECT control Item J to external In this mode the CL 200A will accept external...

Страница 13: ... External Voltage Set BNC Heater Cooler Device 15 pin D connector Temperature Range Peltier 6 C to 65 C Resistive Ambient to 65 C Meter Display Resolution 3 digit LED display of C or V 0 1 C 0 01 Volt 0 01 Ampere Meter Readouts Set temperature C Control Temperature C Monitor Temperature C Output Voltage V Output Current A Enclosure 8 9 X 42 6 X 29 2 cm W x H x D Weight 10 lb 4 54 Kg Operating cond...

Страница 14: ...h a 5x20 mm 0 8 A slow blow fuse 0 4 A for facilities using 220 240 V line voltages If the replacement fuse also fails please call Warner Instruments for assistance B Occasionally a knob on the front panel will loosen after long use These are collet style knobs and are tightened with a screw located under the knob cap To gain access to the adjustment screw pry the cap off with a thin bladed screwd...

Страница 15: ...0000 4 8 EN610000 4 11 Manufacturer s Name Warner Instruments LLC Manufacturer s Address 1125 Dixwell Avenue Hamden CT 06514 Tel 203 776 0664 Equipment Description Power Controller Equipment Class ITE Class A Model Numbers CL 200A I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s Place Hamden Connecticut USA Signature Full Name Bur...

Страница 16: ...ments LLC Manufacturer s Address 1125 Dixwell Avenue Hamden CT 06514 Tel 203 776 0664 Equipment Description Power Controller Safety requirements for electrical equipment for measurement and laboratory use Equipment Class Class I Model Numbers CL 200A I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s Place Hamden Connecticut USA Sig...

Страница 17: ...CL 200A Manual Rev 120523 Warner Instruments A Harvard Apparatus Company 17 ...

Отзывы: