Specifications
Maximum Output Voltage
±15 V Heat/Cool, 0 to 15 V Heat-Only
Maximum Output Current
5.0 Amperes DC per channel
Manual Voltage Range
0 to 15 VDC
Maximum Output Power
75 Watts @ 3
Ω
Load Each Channel
Power Requirements
100-130 or 200-260 VAC, 50/60 Hz, 80 VA
Power Fuse (5 x 20 mm )
0.8 A Slow-Blow for 100-130 VAC
0.4 A Slow-Blow for 200-260 VAC
Front Panel Input
External thermistor
(BNC, calibrated to 10.0 k
at 25°C )
Recorder Outputs
Monitor Temp Out (BNC, 100 mV/°C )
Control Temp Out (BNC, 100 mV/°C )
Rear Panel Inputs
External Temperature Set (BNC, 100 mV/°C )
External Voltage Set (BNC)
Heater/Cooler Device (15-pin D connector)
Temperature Range
Peltier: -6
˚
C to 65
˚
C
Resistive: Ambient to 65
C
Meter Display; Resolution
3 digit LED display of
C or V
0.1
˚
C / 0.01 Volt / 0.01 Ampere
Meter Readouts
Set temperature (°C )
Control Temperature (°C )
Monitor Temperature (°C )
Output Voltage (V )
Output Current (A)
Enclosure
8.9 X 42.6 X 29.2 cm (W x H x D)
Weight
10 lb (4.54 Kg)
Operating conditions:
Ambient temperature: 0 to 40 °C
Altitude: sea level to 2000 m
Relative humidity: 0-95%
This equipment is intended to be operated in a controlled laboratory environment and is not
approved for use with human subjects.
Accessories and replacement parts
Model number
Order number
Description
ACC-1
64-1427
Adapter cable for heater only components
Warner
Instruments
A Harvard Apparatus Company