Maintenance
58
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
CONTENTS
This section deals with the following subjects:
Warnings for maintenance operations
Monthly or 50-cycle maintenance
1200-cycle or yearly maintenance
Warnings for maintenance operations
WARNINGS
WARNING! Turn the sterilizer OFF and remove the power
cord before beginning any maintenance. Follow all health,
safety, cross-infection and cross-contamination
protocols.
Maintenance operation shall be done at illumination level
of 215 lx (±15 lx) to 1500 lx (±15 lx).
Before making any operation, ward off unauthorized
personnel from the working area.
CAUTION!
Before accessing the chamber and the connected
parts, be sure that the sterilizer is cold.
Notice
: follow the instructions in this chapter when carrying out any
maintenance on the sterilizer.
User maintenance
MAINTENANCE BY THE USER
Frequency
1
Cycles
1
Operation
Monthly
50
Cleaning the door gasket and the chamber face
side. See "Cleaning the door gasket and the
chamber face side" on page 60
Clean the chamber, trays and the rack. See
"Cleaning the chamber and the chamber
accessories" on page 61
Cleaning the chamber filters. See "Cleaning the
chamber filters" on page 62
Cleaning the external surfaces of the sterilizer.
See "Cleaning the external surfaces of the
sterilizer" on page 62
Yearly
2
1200
2
Replace the door gasket. See "Replacing the door
gasket" on page 63
Replace the HEPA filter. See "Replacing the HEPA
filter" on page 65
1
: whichever occurs first.