Technical data
84
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Unwrapped
Wrapped Medium
Load
Wrapped Large
Load
Wrapped XL Load
Low Temperature
Sterilization temperature
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
250 °F (121 °C)
Sterilization program type according ANSI/AAMI ST55
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Gravity
displacement
Sterilization program type according EN 13060
Class S
Class S
Class S
Class S
Class S
Sterilization time (minutes)
4'
4'
4'
4'
30'
Drying time (minutes)
Range: 8–99'
Recommended:
8'
Range: 25–99'
Recommended:
25'
Range: 30–99'
Recommended:
30'
Range: 40–99'
Recommended:
40'
Range: 30–99'
Recommended:
30'
Total program duration (min)
1
Full loaded including drying time
26–30'
43–46'
48–52'
58–63'
73–76'
Load type
Pouched, wrapped instruments and dental
handpieces
No
Yes
Yes
Yes
Yes
Unwrapped instruments and dental
handpieces
Yes
Yes
Yes
Yes
Yes
Textiles
No
No
No
No
Yes
Maximum load weight (excluding trays)
instruments:
18 lb (8.1 kg)
instruments:
3 lb (1.4 kg)
instruments:
6 lb (2.7 kg)
instruments:
14 lb (6.3 kg )
included cassettes
instruments:
5 lb (2.2 kg)
included trays
or textile:
4.4 lb (2 kg)
1
: The total cycle time may vary depending on the type of load (solid or porous), the load weight, and other factors.