Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
23
Getting started
Installing the sterilizer
LOCATION REQUIREMENTS
Notice
:
Do not place the sterilizer so that it is difficult to operate the controls
behind the service door. Do not place the sterilizer so that it is
difficult to disconnect the power cord.
Leave the condenser grid (rear side of the sterilizer) free from
anything that might obstruct the air passage.
Surface materials should be water resistant. If sterilization cycles
will be continuous, pay attention to the surrounding materials: steam
can damage them.
The sterilizer must operate in absence of explosive atmospheres.
The sterilizer must operate in a well ventilated room (indoor), far
from sources of heat and from flammable materials.
Place the sterilizer on a flat and level surface.
Clearance requirements to ensure proper air circulation:
ELECTRICAL CONNECTIONS
All the cables and tubes connected on the rear side of the sterilizer
must be placed far from the condenser grid (e.g. using the available
guides).
Notice
:
Connect the sterilizer to a dedicated line. Do not use cable
extensions nor multiple sockets/adapters.
Ensure that external and internal surfaces are free from moisture or
condensation before connecting to power.
The installation of the sterilizer shall be performed by two authorized
technicians using PPE (Personal Protective Equipment) according to
applicable standards.
The electrical power supply of the sterilizer must fulfill all applicable
standards in the country of use, and must comply with the data label
on the back of the sterilizer.
Содержание MN-111
Страница 1: ...MN 111 Instructions for Use MN 111 Med ENG Rev14...
Страница 2: ......