Diagnostics
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
69
Code
Description
Actions
12x
Wait before opening the chamber door. Allow the sterilization chamber to cool down.
Repeat the cycle.
If the problem persists, contact technical service.
Switch the sterilizer OFF and ON.
Clean the chamber and the chamber furniture from residuals of detergents, disinfectants and other
chemicals.
Replace the clean water if it is suspected to be contaminated with chemicals.
Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
13x to 16x
Check water level in the clean water tank. Reset the safety thermostat..
Repeat the cycle.
If the problem persists, contact technical service.
Switch the sterilizer OFF and ON.
Clean the door gasket and the chamber face side.
Check if the load placed in the sterilization chamber complies with the MAXIMUM WEIGHT LIMITS.
Clean the chamber and the chamber furniture from residuals of detergents, disinfectants and other
chemicals.
Replace the clean water if it is suspected to be contaminated with chemicals.
Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
18x
Chamber filters clogged. Remove and clean the chamber filters. See error "13x to 16x" above.
Repeat the cycle.
If the problem persists, contact technical service.
HEPA filter clogged. Check and replace if necessary.
2xx
Switch the sterilizer OFF and ON.
Repeat the cycle.
If the problem persists, contact technical service.
Wait for the chamber to cool down. Reset the safety thermostat (see "Extraordinary maintenance" on
page 66).
3xx
Check the door gasket. Clean or replace it if necessary.
Repeat the cycle.
If the problem persists, contact technical service.
Clean the chamber face side.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.