9032X
32
To abort the operation:
From the Commands window, press the “ABORT OPERATION” button.
3.4.4
DISPLAY WINDOW
The display window will give you access to useful tools for troubleshooting. The display
window will show you the current analyzer reading graphically in mV as well as other
useful information. The analyzer can graph real time the last 10 minutes of the sensor
reading in mV. Each vertical line represents two minutes. The Y axis represents the
sensor output in mV.
To access the Display Window:
Sensor:
Potential reading
Temperature:
Temperature reading of the sample at the flow cell
Offset:
Offset from the strength of the last calibration
Strength:
Percentage of mV1 reading vs mV2 (also referred to as slope)
Flow input:
Displays flow switch output (default ok if no flow switch is installed)