9032X
25
3.2
ALARMS
Sample Concentration Alarms
When the 9032X Sodium system is in normal operation mode one alarm operates as a
“low” alarm and the other operates as a ‘high’ alarm. Alarm 1 (A1) operates as a LOW
alarm and is activated when the sodium level decreases below the set value. Alarm 2
(A2) operates as a HIGH alarm and is activated when the sodium level increases above
the set value. The two sodium alarms control the relays provided. Each relay has one
pair of changeover contacts rated at 5A, 250VAC.
Various Alarm Descriptions
ALARM
DESCRIPTION
CALIBRATION FAIL
Analyzer failed calibration
SAMPLE HOT
Sample temperature over range (131F)
T.SENSOR FAIL
No thermistor response
O/P 1 Out
Concentration is outside the set measurement range for 4-20mA output
O/P 2 Out
Temperature is outside the set measurement range for 4-20mA output
CONC. LOW
Concentration is below Low Alarm set point
CONC. HIGH
Concentration is above High Alarm set point
OVR
Concentration is above limits of analyzer (>10ppm)
OUT
Sample flowrate to analyzer is not sufficient
3.3
GETTING STARTED
o
Insert the POWER cord in the AC socket terminal connector located in
transmitter section. After power-up, the analyzer enters into Measurement
Mode.
o
The analyzer automatically displays the concentration of sodium interpreted by
the electrodes.
o
If the analyzer is being started up for the first time, or if it was not in operation
for a long time, the user should perform a 2 point calibration as detailed in
Section 3.4.3.
3.4
SOFTWARE PROGRAMMING
The X Module Analyzer has been designed to retain all the same functionality as
the superseded Model 9032, but with a much more user friendly interface. It is
equipped with a touch screen HMI that allows more straightforward navigation
through the menu options as well as enhanced capabilities. Improvements
include increase data log storage space, programming to avoid the high
concentration alarm after running calibration, and options to add different
languages.