Enter in the expected concentration of the sample and press “ENT.” You will then
see the entered value in the “Process cal. value” window. If you wish to make
changes to the entered value, press the number “XX.X ppb” to edit the value and hit
“ENT” again. Press “CALIBRATE” when ready. The Process Calibration will begin.
3.4.3.5
ABORT OPERATION
Any calibration cycle or Grab Sample cycle may be interrupted at any time by pressing
the ABORT OPERATION button. Once the calibration process is aborted, the
measurement screen is displayed.