16
NOTE: Disinfectant concentration can change rapidly in the sample!
Minimize the time between performing the DPD test or manual analysis and
finishing the calibration!
4.0 Troubleshooting
The disinfectant reading is much lower than the manual
analysis
Possible Causes
Corrective Actions
Insufficient conditioning
Wait for the appropriate amount of time
before attempting a calibration
Insufficient sample flow
Increase flow rate to between 20 and
100
liter per hour
Air bubbles in sample
Dislodge bubbles
Adjust flow rate higher if necessary
Allow sample to outgas
Loose electrolyte housing
Tighten electrolyte housing
Protection cap not removed
Remove protection cap
Dirty electrodes
Clean electrodes with emery paper
Replace balls in cleaning attachment
No electrolyte fill solution in
electrolyte housing
Fill electrolyte housing with electrolyte
Air bubbles in electrolyte
Empty electrolyte housing and refill it
Faulty cable
Replace cable
Faulty sensor
Replace sensor
Faulty analysis equipment or
reagents
Consult test equipment instructions
The disinfectant reading is much higher than the manual
analysis
Possible Causes
Corrective Actions
Insufficient conditioning
Wait for the appropriate amount of time
before attempting a calibration
Faulty sensor
Replace sensor
Faulty analysis equipment or
reagents
Consult test equipment instructions
Sample contaminated with
interfering molecule (refer to
Sensitivity specification in
Section 6)
Remove source of contamination