3.
SLOWLY screw on the housing until it is hand tight.
Be prepared
for some electrolyte solution to squeeze out.
4.
Rinse your hands, the sensor, and all surfaces contaminated with
electrolyte solution with running water. Check the sensor for leaks.
5.
Push the cable onto the end of the sensor, aligning the pins with the
holes. Turn the connector until hand tight to seal the cable
connection.
Sensors with optional cleaning attachment
The cleaning attachment consists of a cleaning chamber and two
bags of glass balls.
When using the cleaning attachment, the
minimum sample flow rate is 45 l/hour!
The cleaning attachment
is only effective in removing thin deposits.
1.
Complete the steps above for sensors without cleaning attachment.
2.
Remove the o-ring holder from the acrylic flow cell and slide it up
onto the sensor to sit below the o-ring.
3.
Place three of the glass balls into the cleaning chamber and screw
it onto the electrolyte housing.