2.0 Installation
Assembling the Sensor
CAUTION:
Wear gloves and safety glasses during assembly of
the sensor since the electrolyte is a diluted acid. It is recommended
to perform this operation over a sink with running water available.
Please heed the warnings on the electrolyte bottle. Do not swallow
the electrolyte. Avoid contact of the electrolyte with skin and eyes.
Otherwise wash with a lot of water. In case of eye inflammation,
contact a doctor.
After using, re-cap any remaining electrolyte and store the bottle
upside-down until the next use.
Never shake the electrolyte bottle, as this will introduce air bubbles
that will negatively impact performance!
Do not touch or otherwise contaminate the electrodes!
For PEEK sensors, never remove the reservoir cartridge
attached to the reference electrode!
Sensors without optional cleaning attachment
1.
Holding the sensor at the housing as shown, unscrew the dark grey
protective cap. The cap contains liquid. Save the protective cap in
case the electrode will need to be stored for more than a month of
downtime.
2.
Unscrew the housing until there is a gap, then fill it to the top with
the electrolyte until it overflows.
Never shake the electrolyte
bottle, it must stay free of bubbles!