20
heed the warnings on the electrolyte bottle. Do not swallow the
electrolyte. Avoid contact of the electrolyte with skin and eyes.
Otherwise wash with a lot of water. In case of eye inflammation,
contact a doctor.
To store the sensor unscrew the electrolyte housing.
Rinse the electrolyte housing and the electrode finger in clean water
and dry in a dust-free location.
Replace the protective cap onto the electrolyte housing.
Loosely screw the electrolyte housing with protective cap onto the
electrode shaft to protect the electrodes.
When putting the sensor back into operation
after storage, the
electrodes must be cleaned with the emery paper.