For each of the protocols listed you can perform the following actions. Simply select the  protocol 

and click on one of the buttons: 

• View:

 

view a protocol. No parameters can be changed in the “View mode”. 

•  Edit:

  edit  a  protocol  (change  the  VWR  Electronic  Pipette  model,  rename  protocol,  add/edit/ 

remove tasks). 

• Clone:

 clone a protocol (if you are creating a similar protocol, cloning will save time). 

• Export:

 export a protocol to a .pmp file and store it on the computer. 

• Export to pdf:

 create a summary of the protocol in the form of a print-ready pdf document. 

• Delete:

 delete a protocol. 

Alternatively, you can perform these actions by right-clicking on a desired protocol. 

Editing protocol is also available by double-clicking the selected protocol. 

 

 

5.3 Importing a protocol: 

You can add protocols to the 

Manage custom protocols 

by importing a previously created 

protocol (.pmp files). 

Click on 

Import a protocol from file 

button, navigate to the .pmp file and click 

Open

The protocol is then available in 

Manage custom protocols

 

 

5.4 Practical tips for creating custom protocols: 

If you have existing protocols, cloning and modifying them will save you time. 

VWR Protocol Builder Software enables you to easily create a similar protocol on a different 

VWR Electronic Pipette model. Simply clone the protocol and change the VWR Electronic Pipette 

model.  If  necessary,  modify  the  volume  of  transfers  to  fit  volume  specifications  of  the  new  VWR 

Electronic Pipette model. 

 

6

 Transferring custom protocols to the pipette 

To transfer a protocol to a VWR Electronic pipette, click the 

Transfer custom protocols to pipette 

button and follow the instructions on the screen: 

1/  Connect  pipette  to  the  PC  via  USB  cable  and  click 

Next

Note:

  should  the  pipette  not  be 

recognized  immediately,  disconnect  the  USB  cable,  reconnect  it  and  click 

Try  again

  button.  If  a 

pipette is already correctly connected when 

Transfer custom protocols to pipette

 button is clicked, 

this step is skipped. 

2/ By default, the protocols for the connected VWR Electronic Pipette model are listed. You can 

access the protocols created for other VWR Electronic Pipette models by selecting the model from 

the 

Show protocols for pipette model

(2)

: drop-down list. 

3/ Select a protocol in the list and click 

Finish. 

4/  If  a  pipette  is  locked,  transferring  protocols  will  unlock  it.  For  that  purpose,  a  pop-up  window 

appears, where you can confirm that you would like to transfer the selected protocol, unlocking the 

pipette in the process. 

Содержание VWR 10

Страница 1: ...0 L 10827 934 VWR ELECTRONIC PIPETTE 12X20 L 10827 936 VWR ELECTRONIC PIPETTE 12X300 L 10827 938 VWR ELECTRONIC PIPETTE 12X1200 L European Catalog Numbers 613 5415 VWR ELECTRONIC PIPETTE 10 L 613 5416...

Страница 2: ...85011 http be vwr com Country of origin FRANCE Intended use VWR ELECTRONIC PIPETTE is intended to be used in liquid handling applications for dispensing liquids in volume range of 0 5 l to 1200 l This...

Страница 3: ...RY PRACTICE FEATURES 17 11 VWR ELECTRONIC PIPETTE POWER MANAGEMENT 17 12 CONFIGURATION 19 13 MAINTENANCE OF YOUR VWR ELECTRONIC PIPETTE 20 14 CLEANING AND DECONTAMINATION 25 15 LEAK TEST 27 16 TROUBLE...

Страница 4: ...I Repetitive strain injury associated with routine work It combines superior accuracy precision and robustness with advanced ergonomics design it fits comfortably in either left or right hand and allo...

Страница 5: ...0827 940 613 5427 VWR CHARGING STAND ADAPTER WHITE 10827 942 613 5428 US Europe VWR 10 10827 884 613 5415 0 5 10 l Red VWR 20 10827 918 613 5416 2 20 l Pale Yellow VWR 300 10827 920 613 5417 20 300 l...

Страница 6: ...utton to reactivate the pipette your last setting and pipetting mode will appear on the screen 4 DESCRIPTION VWR Electronic Pipette single and multichannel models Upper part VWR Electronic Pipette is...

Страница 7: ...er Optimized tip holder to reduce tip fitting and ejection forces removable for cleaning and servicing Ejector clip Ejector support Cover Ejector spacer only for Multi 1200 l Pipetting Mode Aspirate a...

Страница 8: ...click to change volume Now press the Push button the volume on the display will start blinking you can now adjust the volume Turn the Push button either clockwise to decrease volume or counter clockwi...

Страница 9: ...while maintaining the volume and the last settings Continue pipetting as you would as if you were using a mechanical pipette REPETITIVE Mode The REPETITIVE Mode allows you to distribute the same volu...

Страница 10: ...ired extra volume The extra volume is necessary to ensure equal operating conditions for each dispensed aliquot 3 To dispense click the Push button DISCARD appears on the screen Discard the extra volu...

Страница 11: ...press the push button Set the additional pipetting volume named VOLUME 3 You can choose a volume to be aspirated and dispensed after the mixing step so you can continue routine pipetting Click the Pus...

Страница 12: ...sh button set the volume by turning the Push button and click again to confirm your settings You are ready to do reverse pipetting 1 Fit a suitable tip 2 Click the Push button to aspirate the selected...

Страница 13: ...of the computer if Create desktop shortcut is checked 4 If VWR Protocol builder software is not launched automatically double click the VWR Protocol builder icon on the desktop 5 Once installed the V...

Страница 14: ...by turning the Push button and click to confirm your selection Set DISPENSING SPEED dispensing speed is blinking set the speed by turning the Push button and click to confirm your selection After sele...

Страница 15: ...e down Never lay the pipette on its side when there is liquid in the tip s g The VWR charging Stand Adapter refer to chapter 2 is recommended for use with the VWR Carrousel Pipette Stand to store the...

Страница 16: ...nsion for the Single Channel Hold the pipette with the push button upright Hold the extension with the slot upright Slide the extension over the tip holder Push the extension firmly onto the end of th...

Страница 17: ...Bar code on the box Personalization by nametags for marking application and or user name refer to chapter 9 Cycle counters From last volume setting to count the number of cycles in the current run Fr...

Страница 18: ...ette The pipette starts charging 2 Charge your pipette by using VWR Charging stand adapter First place the stand adapter on your VWR Carrousel Pipette Stand Remove the plug Now place your pipette on t...

Страница 19: ...ry provides you very short cycle time or permanently needs to be used in connected mode 12 CONFIGURATION Enter the menu using the mode button You will find not only the pipetting programs PIPET REPETI...

Страница 20: ...or autoclaving by disassembly of the parts specified refer to chapter 13 Service information VWR Electronic Pipette provides you with all required service information It will help you establish an ea...

Страница 21: ...u to access the following items Standard readjustment Reset settings Standard readjustment This menu allows you to adjust the pipette using 3 calibration points 10 50 and 100 of nominal volume in acco...

Страница 22: ...ly switch to disassembly mode so you can carry out your maintenance safely Wait longer than 10 seconds and the option to have a reminder in one week will appear 3 Click to confirm Your pipette will be...

Страница 23: ...e that the two parts are fully tightened by hand 6 Refit the tip ejector After autoclaving the tip holder and tip ejector may change color however this has no impact on the performance Changing the se...

Страница 24: ...assemble the piston and guide then carefully insert the assembly into the body of the pipette The guide should hold the piston assembly inside the body of the pipette You can hear a click when the pis...

Страница 25: ...act with contaminants can easily be cleaned and decontaminated If you use chemical solutions for decontamination or detergents for cleaning other than those specified below you should check with your...

Страница 26: ...the cleaning solution or wipe them with a soft cloth or lint free tissue impregnated with the cleaning solution 3 Rinse the components with distilled water 4 Leave the parts to dry by evaporation or...

Страница 27: ...2 Immerse the components in the decontaminant solution or wipe them according the instructions given by the manufacturer or supplier of the decontaminant 3 Rinse the individual components with distill...

Страница 28: ...consult the following table which identifies potential problems and their solutions 17 SPECIFICATIONS VWR Electronic Pipette is a high quality pipette that offers excellent accuracy and precision it...

Страница 29: ...CAUTIONS AND LIMITATIONS OF USE For safety reasons it is important to observe the following instructions 1 Battery and electrical specifications Use AC adaptor and stand adapter indoors VWR electronic...

Страница 30: ...gles and gloves and regulations appropriate for your country 4 Do not allow the liquid to enter the body of the pipette 5 Storage conditions Temperature 20 C to 50 C Humidity max 80 6 Temperature of u...

Страница 31: ...31 19 REPLACEMENT PARTS COMING SOON...

Страница 32: ...32 APPENDIX 1 EXAMPLE OF A PERFORMANCE CHECK example with VWR10 at 1 l...

Страница 33: ...33 APPENDIX 2 Z FACTOR...

Страница 34: ...a sample 4 Tare the balance and take the weighing vessel out of the balance 5 Take off the lid with tweezers 6 Dispense the sample into a dummy vessel 7 Replace the lid on the weighing vessel and usin...

Страница 35: ...nstalling VWR Protocol Builder Software 5 Creating and managing custom protocols 5 1 Creating a custom protocol 5 1 1 Default task parameters 5 1 2 Editing cloning and deleting tasks 5 1 3 Pipette vol...

Страница 36: ...hange password User profile Provide Institution Company details Providing this information is optional and can also be done later in main menu select Other administration Institution Company profile D...

Страница 37: ...e and transfer a selected custom protocol to the VWR Electronic pipette Update pipette settings Access additional pipette settings such as locking functionality sound alerts maximum volume limit or pi...

Страница 38: ...cked iv If VWR Protocol Builder Software is not launched automatically double click the VWR Protocol Builder Software icon on the desktop v Once installed the VWR Protocol Builder Software can be run...

Страница 39: ...creen to the Protocol panel on the right of the screen and modifying the task s default parameters 1 Select a task in the Available task list 2 Drag and drop it to the Protocol panel use the arrow but...

Страница 40: ...seconds Beep makes pipette produce a beeping sound while selected tasks are executed Add tasks to the protocol through drag and drop actions by selecting a task and clicking the arrow button or by do...

Страница 41: ...er of characters for display text is limited 3 Dispense all task has the following parameters Task name 2 specify the name of the task Speed 2 specify the speed for the dispense Comments enter any com...

Страница 42: ...ait task has the following parameters Task name 2 specify the name of the task Duration s 2 specify the duration of the pause in seconds Comments enter any comments related to the task Display text 2...

Страница 43: ...u can select a task in the protocol and click the left arrow When deleting a task a pop up window appears where you can confirm task deletion 5 1 3 Pipette volume indicator As you build a protocol the...

Страница 44: ...e and cannot be exported In certain cases a yellow warning sign will appear next to a protocol in Manage custom protocols and Transfer custom protocols to pipette More details about the warning are av...

Страница 45: ...ou time VWR Protocol Builder Software enables you to easily create a similar protocol on a different VWR Electronic Pipette model Simply clone the protocol and change the VWR Electronic Pipette model...

Страница 46: ...tered in the software prior to transferring custom protocols If the connected pipette is not yet registered a pop up appears which informs you that the pipette is not yet registered Here you can click...

Страница 47: ...volume on the particular pipette it cannot be higher than pipette s specified maximal volume Simply enter the volume in L in the field Update Firmware To modify these configurations follow these steps...

Страница 48: ...Updating pipette firmware To update firmware VWR Electronic pipette follow these steps 1 Click Update pipette settings in the main menu 2 If a pipette is already correctly connected proceed to step 4...

Страница 49: ...et the maintenance dates last and next by clicking the Refresh dates button This process requires you to connect the selected pipette to the computer You can delete a pipette in the Registered pipette...

Страница 50: ...tab select the time interval and click Show history The protocols that have been modified within selected time interval will be listed in the table below displaying the following information for each...

Страница 51: ...sswords checkbox Don t use password Note remember or store the answer to the hidden question in case of password recovery 8 4 Institution Company profile Enter the information related to your Institut...

Страница 52: ...52...

Страница 53: ...epair replace or refund the purchase price of this product to the customer provided it is returned during the warranty period This warranty does not apply if the product has been damaged by accident a...

Страница 54: ...r responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the disposal and recycling...

Страница 55: ...vwr_india vwr com Ireland Northern Ireland VWR International Ltd VWR International Northern Ireland Ltd Orion Business Campus Northwest Business Park Ballycoolin Dublin 15 Tel 01 88 22 222 Fax 01 88 2...

Отзывы: