36 

 

1

 INTRODUCTION 

 

1.1 What is VWR Protocol Builder Software? 

VWR Protocol Builder Software is an easy-to-use software for creating custom pipetting protocols 

for  VWR  Electronic  Pipettes.  Use  it  to  create  personalized  pipetting  protocols  on  your  computer, 

and to transfer these protocols to  VWR Electronic Pipettes in the 

Custom Mode

, where you can 

execute them. 

VWR Protocol Builder Software runs on Windows 7, Windows 8 and Windows 8.1. 

The software file is downloadable on 

www.vwr.com/e-pipette

.

 

 

 

1.2 Passwords and User information 

When running the VWR Protocol Builder Software for the first time (e.g. after installation), you are 

prompted to enter some information regarding your password and your institution. 

 

User password: 

Enter  user  password  and  select  a 

Hidden  question  /  Hidden  answer 

(used  to  recover  forgotten 

passwords). You can also select 

I don’t want to use passwords.

 

You  can  still  switch  to  start  using  passwords  later  (in  main  menu  select 

Other  administration 

► 

Change password)

 

User profile: 

Provide  Institution/Company  details.  Providing  this  information  is  optional  and  can  also  be  done 

later (in main menu select 

Other administration 

Institution/Company profile

). 

During the installation you are prompted to enter a 

Recovery password 

that will be used to retrieve 

forgotten passwords. 

Password must contain 8-30 characters, start with a letter, contain at least one numerical 

digit and one uppercase letter (no special symbols allowed). 

Note:

 When using a password, please make sure you remember or save your 

Recovery password 

and/or the answer to the 

Hidden question 

– those are the only two means of retrieving your login 

information in the event of forgetting the password (see following section). 

 

 

1.3 Forgotten password 

Users who use password protection have to provide the password to run VWR Protocol Builder 

Software and to modify custom protocols in VWR Electronic Pipette. 

 

In the case of a forgotten password, click the 

Forgotten Password 

button on the  VWR Electronic 

Pipette login window. Select one of the following options to recover the forgotten password: 

• I know the answer to the hidden question 
• I have the recovery password 

 

Содержание VWR 10

Страница 1: ...0 L 10827 934 VWR ELECTRONIC PIPETTE 12X20 L 10827 936 VWR ELECTRONIC PIPETTE 12X300 L 10827 938 VWR ELECTRONIC PIPETTE 12X1200 L European Catalog Numbers 613 5415 VWR ELECTRONIC PIPETTE 10 L 613 5416...

Страница 2: ...85011 http be vwr com Country of origin FRANCE Intended use VWR ELECTRONIC PIPETTE is intended to be used in liquid handling applications for dispensing liquids in volume range of 0 5 l to 1200 l This...

Страница 3: ...RY PRACTICE FEATURES 17 11 VWR ELECTRONIC PIPETTE POWER MANAGEMENT 17 12 CONFIGURATION 19 13 MAINTENANCE OF YOUR VWR ELECTRONIC PIPETTE 20 14 CLEANING AND DECONTAMINATION 25 15 LEAK TEST 27 16 TROUBLE...

Страница 4: ...I Repetitive strain injury associated with routine work It combines superior accuracy precision and robustness with advanced ergonomics design it fits comfortably in either left or right hand and allo...

Страница 5: ...0827 940 613 5427 VWR CHARGING STAND ADAPTER WHITE 10827 942 613 5428 US Europe VWR 10 10827 884 613 5415 0 5 10 l Red VWR 20 10827 918 613 5416 2 20 l Pale Yellow VWR 300 10827 920 613 5417 20 300 l...

Страница 6: ...utton to reactivate the pipette your last setting and pipetting mode will appear on the screen 4 DESCRIPTION VWR Electronic Pipette single and multichannel models Upper part VWR Electronic Pipette is...

Страница 7: ...er Optimized tip holder to reduce tip fitting and ejection forces removable for cleaning and servicing Ejector clip Ejector support Cover Ejector spacer only for Multi 1200 l Pipetting Mode Aspirate a...

Страница 8: ...click to change volume Now press the Push button the volume on the display will start blinking you can now adjust the volume Turn the Push button either clockwise to decrease volume or counter clockwi...

Страница 9: ...while maintaining the volume and the last settings Continue pipetting as you would as if you were using a mechanical pipette REPETITIVE Mode The REPETITIVE Mode allows you to distribute the same volu...

Страница 10: ...ired extra volume The extra volume is necessary to ensure equal operating conditions for each dispensed aliquot 3 To dispense click the Push button DISCARD appears on the screen Discard the extra volu...

Страница 11: ...press the push button Set the additional pipetting volume named VOLUME 3 You can choose a volume to be aspirated and dispensed after the mixing step so you can continue routine pipetting Click the Pus...

Страница 12: ...sh button set the volume by turning the Push button and click again to confirm your settings You are ready to do reverse pipetting 1 Fit a suitable tip 2 Click the Push button to aspirate the selected...

Страница 13: ...of the computer if Create desktop shortcut is checked 4 If VWR Protocol builder software is not launched automatically double click the VWR Protocol builder icon on the desktop 5 Once installed the V...

Страница 14: ...by turning the Push button and click to confirm your selection Set DISPENSING SPEED dispensing speed is blinking set the speed by turning the Push button and click to confirm your selection After sele...

Страница 15: ...e down Never lay the pipette on its side when there is liquid in the tip s g The VWR charging Stand Adapter refer to chapter 2 is recommended for use with the VWR Carrousel Pipette Stand to store the...

Страница 16: ...nsion for the Single Channel Hold the pipette with the push button upright Hold the extension with the slot upright Slide the extension over the tip holder Push the extension firmly onto the end of th...

Страница 17: ...Bar code on the box Personalization by nametags for marking application and or user name refer to chapter 9 Cycle counters From last volume setting to count the number of cycles in the current run Fr...

Страница 18: ...ette The pipette starts charging 2 Charge your pipette by using VWR Charging stand adapter First place the stand adapter on your VWR Carrousel Pipette Stand Remove the plug Now place your pipette on t...

Страница 19: ...ry provides you very short cycle time or permanently needs to be used in connected mode 12 CONFIGURATION Enter the menu using the mode button You will find not only the pipetting programs PIPET REPETI...

Страница 20: ...or autoclaving by disassembly of the parts specified refer to chapter 13 Service information VWR Electronic Pipette provides you with all required service information It will help you establish an ea...

Страница 21: ...u to access the following items Standard readjustment Reset settings Standard readjustment This menu allows you to adjust the pipette using 3 calibration points 10 50 and 100 of nominal volume in acco...

Страница 22: ...ly switch to disassembly mode so you can carry out your maintenance safely Wait longer than 10 seconds and the option to have a reminder in one week will appear 3 Click to confirm Your pipette will be...

Страница 23: ...e that the two parts are fully tightened by hand 6 Refit the tip ejector After autoclaving the tip holder and tip ejector may change color however this has no impact on the performance Changing the se...

Страница 24: ...assemble the piston and guide then carefully insert the assembly into the body of the pipette The guide should hold the piston assembly inside the body of the pipette You can hear a click when the pis...

Страница 25: ...act with contaminants can easily be cleaned and decontaminated If you use chemical solutions for decontamination or detergents for cleaning other than those specified below you should check with your...

Страница 26: ...the cleaning solution or wipe them with a soft cloth or lint free tissue impregnated with the cleaning solution 3 Rinse the components with distilled water 4 Leave the parts to dry by evaporation or...

Страница 27: ...2 Immerse the components in the decontaminant solution or wipe them according the instructions given by the manufacturer or supplier of the decontaminant 3 Rinse the individual components with distill...

Страница 28: ...consult the following table which identifies potential problems and their solutions 17 SPECIFICATIONS VWR Electronic Pipette is a high quality pipette that offers excellent accuracy and precision it...

Страница 29: ...CAUTIONS AND LIMITATIONS OF USE For safety reasons it is important to observe the following instructions 1 Battery and electrical specifications Use AC adaptor and stand adapter indoors VWR electronic...

Страница 30: ...gles and gloves and regulations appropriate for your country 4 Do not allow the liquid to enter the body of the pipette 5 Storage conditions Temperature 20 C to 50 C Humidity max 80 6 Temperature of u...

Страница 31: ...31 19 REPLACEMENT PARTS COMING SOON...

Страница 32: ...32 APPENDIX 1 EXAMPLE OF A PERFORMANCE CHECK example with VWR10 at 1 l...

Страница 33: ...33 APPENDIX 2 Z FACTOR...

Страница 34: ...a sample 4 Tare the balance and take the weighing vessel out of the balance 5 Take off the lid with tweezers 6 Dispense the sample into a dummy vessel 7 Replace the lid on the weighing vessel and usin...

Страница 35: ...nstalling VWR Protocol Builder Software 5 Creating and managing custom protocols 5 1 Creating a custom protocol 5 1 1 Default task parameters 5 1 2 Editing cloning and deleting tasks 5 1 3 Pipette vol...

Страница 36: ...hange password User profile Provide Institution Company details Providing this information is optional and can also be done later in main menu select Other administration Institution Company profile D...

Страница 37: ...e and transfer a selected custom protocol to the VWR Electronic pipette Update pipette settings Access additional pipette settings such as locking functionality sound alerts maximum volume limit or pi...

Страница 38: ...cked iv If VWR Protocol Builder Software is not launched automatically double click the VWR Protocol Builder Software icon on the desktop v Once installed the VWR Protocol Builder Software can be run...

Страница 39: ...creen to the Protocol panel on the right of the screen and modifying the task s default parameters 1 Select a task in the Available task list 2 Drag and drop it to the Protocol panel use the arrow but...

Страница 40: ...seconds Beep makes pipette produce a beeping sound while selected tasks are executed Add tasks to the protocol through drag and drop actions by selecting a task and clicking the arrow button or by do...

Страница 41: ...er of characters for display text is limited 3 Dispense all task has the following parameters Task name 2 specify the name of the task Speed 2 specify the speed for the dispense Comments enter any com...

Страница 42: ...ait task has the following parameters Task name 2 specify the name of the task Duration s 2 specify the duration of the pause in seconds Comments enter any comments related to the task Display text 2...

Страница 43: ...u can select a task in the protocol and click the left arrow When deleting a task a pop up window appears where you can confirm task deletion 5 1 3 Pipette volume indicator As you build a protocol the...

Страница 44: ...e and cannot be exported In certain cases a yellow warning sign will appear next to a protocol in Manage custom protocols and Transfer custom protocols to pipette More details about the warning are av...

Страница 45: ...ou time VWR Protocol Builder Software enables you to easily create a similar protocol on a different VWR Electronic Pipette model Simply clone the protocol and change the VWR Electronic Pipette model...

Страница 46: ...tered in the software prior to transferring custom protocols If the connected pipette is not yet registered a pop up appears which informs you that the pipette is not yet registered Here you can click...

Страница 47: ...volume on the particular pipette it cannot be higher than pipette s specified maximal volume Simply enter the volume in L in the field Update Firmware To modify these configurations follow these steps...

Страница 48: ...Updating pipette firmware To update firmware VWR Electronic pipette follow these steps 1 Click Update pipette settings in the main menu 2 If a pipette is already correctly connected proceed to step 4...

Страница 49: ...et the maintenance dates last and next by clicking the Refresh dates button This process requires you to connect the selected pipette to the computer You can delete a pipette in the Registered pipette...

Страница 50: ...tab select the time interval and click Show history The protocols that have been modified within selected time interval will be listed in the table below displaying the following information for each...

Страница 51: ...sswords checkbox Don t use password Note remember or store the answer to the hidden question in case of password recovery 8 4 Institution Company profile Enter the information related to your Institut...

Страница 52: ...52...

Страница 53: ...epair replace or refund the purchase price of this product to the customer provided it is returned during the warranty period This warranty does not apply if the product has been damaged by accident a...

Страница 54: ...r responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the disposal and recycling...

Страница 55: ...vwr_india vwr com Ireland Northern Ireland VWR International Ltd VWR International Northern Ireland Ltd Orion Business Campus Northwest Business Park Ballycoolin Dublin 15 Tel 01 88 22 222 Fax 01 88 2...

Отзывы: