11
MIX Mode
This is the classic pipette mode (forward pipetting) followed by a mixing phase, composed of
repeatedly aspirating and dispensing, as well as an optional forward pipetting step.
• Press the Mode-button to access menu, where you can choose between different pipette modes.
Select “
MIX Mode
” by turning the Push-button, to confirm your selection click on the Push-button.
• Set the first volume, named “
VOLUME 1
”
: click the Push-button, set the volume by turning the
Push-button and click to confirm your selection. This is the first volume you will aspirate and
dispense.
Set the second volume, named
“
VOLUME 2
”
: click the Push-button, you can now set the second
volume by turning the Push-button. Click again to confirm your settings.
If you set
“VOLUME 1”
or
“VOLUME 2”
= 0
μL, the step will not appear. If you set
“VOLUME 1”
and
“VOLUME 2”
= 0
μL the cycle will start directly with the MIX step.
• Set the
MIX
volume: click the Push-button, set the volume by turning the Push-button. Confirm
your settings with a simple click. This is the volume which will be aspirated and dispensed
repeatedly.
“MIX” as long as you press the push-button.
• Set the additional pipetting volume, named
“VOLUME 3”
: You can choose a volume to be
aspirated and dispensed after the mixing step, so you can continue routine pipetting.
Click the Push-button, set the volume by turning the Push-button. Confirm your settings by a
simple click.
If you set the
“VOLUME 3 “
= 0
μL the cycle will stop after the
MIX
step.
You are ready to mix!
1/ Fit a suitable tip
2/ To aspirate
“VOLUME 1”
press the Push-button
3/ To dispense: Click (press and release) the Push-button: three things happen: sample dispense,
automatic purge and piston reset to zero
or
Keep your thumb pressing the Push-button until the end of the dispense cycle: the piston reset
takes place after you release the Push-button again. This feature enables you to remove the tip
from the liquid, without aspirating anything.
4/ To aspirate and dispense the second volume
“VOLUME 2”
, repeat steps 2 and 3.
5/ To mix: press the Push-button; as long as you keep your thumb pressing the button, the pipette
continues mixing. Release the Push-button to complete the current mixing cycle.
6/ To purge: press push-button,
“PURGE”
appears on the screen, click again to purge and reset
the piston.
7/ To pipette: if you have chosen a
“VOLUME 3”
> 0, you can now aspirate and dispense this
volume by repeating steps 2 and 3.
Содержание VWR 10
Страница 31: ...31 19 REPLACEMENT PARTS COMING SOON...
Страница 32: ...32 APPENDIX 1 EXAMPLE OF A PERFORMANCE CHECK example with VWR10 at 1 l...
Страница 33: ...33 APPENDIX 2 Z FACTOR...
Страница 52: ...52...