Assembly instructions VOSSOBW501
SG sensor
, Page 13 of 22
Step 3
Align the sensor and prefix the cable with cable ties.
Do not completely fix the cable ties for final aligning
later (see step 7).
Do not contaminate the cleaned surface and the
sensor.
Fig. 23: Aligning the sensor and prefixing the cable ties
Step 4
Prepare the adhesive: use the shown pieces on the right
(Fig. 24): Panacol GTH-T with mixer tip and two-part
adhesive applicator.
Use only one mixer tip for one sensor assembly. For
two sensors only if adhesive is added to both sensors
in less than one minute. Remember to order extra
mixer tips if needed.
The storage temperature of the adhesive must be in
accordance with the manufacturer's
recommendations.
Fig. 24: Panacol GTH-T with mixer tip and two-part adhesive
applicator
Step 4.1
Prepare the adhesive: keep the cartridge in an upright
position for at least 2 minutes, then remove the cap and
attach the mixer tip.
Fig. 25: Preparing the adhesive: attaching mixer tip
!
!
!
!