Indication of the accuracy in ± (% of the full measuring range). The accuracy is valid for one year at a temperature
of +23 °C ± 5 °C, and at a relative humidity of less than 80%, non-condensing. Additional temperature coefficient for
<18 °C and >28 °C: 0.2 x (specified accuracy)/1 °C
Alternate current (AC-A)
Measuring range
Resolution
Accuracy
3 - 30 A/AC
100 mV/A
±(3% of FS) for 45Hz - 500Hz
30 - 300 A/AC
10 mV/A
300 - 3000 A/AC
1 mV/A
Noise level at the output: < 8 mV for all measuring ranges
The accuracies apply to 10% to 100 % of the measuring ranges.
Additional error limits at deviation of the current-conducting conductor from the centre of the Flex current sensor:
VC-10T
VC-18T
Distance
Error
Distance
Error
15 mm
±2%
35 mm
±1%
25 mm
±2.5%
50 mm
±1.5%
35 mm
±3%
60 mm
±2%
do not exceed the maximum permitted input values. do not touch any circuits or parts of circuits
if they may be subject to voltages higher than 33 v/ACrms or 70 v/dC! danger to life!