64
Direct voltage V/DC
Range
Resolution
Accuracy
600.0 mV*
0.1 mV
±(0.9% + 6)
6.000 V
0.001 V
±(0.7% + 6)
60.00 V
0.01 V
±(0.9% + 4)
600.0 V
0.1 V
1000 V
1 V
* only available via the measuring function “mV”
Overload protection 1000 V; Impedance: 10 MΩ (mV: ≥100 MΩ)
When the measuring input is short-circuited in the mV area, a display of 5 counts is possible
Alternating voltage V/AC
Range
Resolution
Accuracy
600.0 mV*
0.1 mV
±(1.3% + 5)
6.000 V
0.001 V
±(1.0% + 5)
60.00 V
0.01 V
±(1.3% + 4)
600.0 V
0.1 V
750 V
1 V
600.0 V “LoZ”
0.1 V
±(2.6%+4)
600.0 V “Motor”**
0.1 V
±(2.0% + 6)
* only available via the measuring function “mV”
** the precision indicated is not valid for speed-controlled motor drives
Frequency range 40 - 400 Hz; Overload protection 750 V; Impedance: 10 MΩ (mV: ≥100 MΩ)
Specified measurement range: 5 - 100 % of the measuring range
When the measuring input is short-circuited, a display of 5 counts is possible
After use of the LoZ function, a regeneration time of 1 minute is required
TrueRMS crest factor (CF) ≤3 CF to 600 V, ≤1.5 CF to 750 V
TrueRMS crest factor for non-sine-shaped signals plus tolerance addition:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Criteria for phase recognition variable frequency in the “motor” mode: >80 V/AC, from 50 to 80 Hz