![Vogtlin SmartTrak 100 Series Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/vogtlin/smarttrak-100-series/smarttrak-100-series_manual_1044498049.webp)
49
1.
Top Level Screens that display information In our software, these screens appear as
yellow boxes. They include:
Mass flow rate
Gas (10 options pre-programmed, one must be AIR)
Engineering units (mass per unit time)
Current Setpoint with units
Source of Setpoint (analog or digital and type)
Valve operation mode (normal, valve shut or purge)
Current meter full scale value with units (user selectable)
2.
Lower level Screens that permit changes to instrument operation In our software, these
screens appear as white boxes. They include:
Setpoint value
Engineering units
Gas
Valve operation
Source of the setpoint signal
Form of the output signals
Full scale of the instrument
Zero meter
Span meter
3.
Additional Features include:
Re-boot the SmartTrak microprocessor
Change the Communication Port
Links to Vögtlin Instruments’ Web Site
Specifications
Power Up Your Instrument
CAUTION: The SmartTrak is not a loop-powered device. Do not apply power to the
0/4-20mA outputs.
After your instrument is installed and the system has undergone a complete leak check
(discussed in detail in Chapter 2), apply power using Vögtlin’s power supply or your own
input power source. See Chapter 2, for power supply requirements. The green LED at the
top of the left side will light. If your instrument has a Pilot Module, it will begin its start-
up cycle. See Chapter 4 for details on Pilot Module operation.
If you have a Mass Flow Controller, the valve will remain closed until power is supplied.
Remember that the valve in the SmartTrak controller is not a positive shut-off device.
Содержание SmartTrak 100 Series
Страница 63: ...63 C ADJUSTMENTS Clicking on adjustments brings up the meter settings screen ...
Страница 74: ...72 Appendix A C100 M100 SmartTrak Product Specifications Dimensions and Mounting ...
Страница 75: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 73 ...
Страница 76: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 4 7 ...
Страница 77: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 75 ...
Страница 78: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 76 ...
Страница 79: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 77 ...
Страница 80: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 78 ...
Страница 82: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 80 ...
Страница 83: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 81 ...
Страница 85: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 83 ...
Страница 86: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 84 ...
Страница 87: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 85 ...
Страница 89: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 87 ...
Страница 90: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 88 ...
Страница 91: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 89 ...
Страница 92: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 90 ...
Страница 109: ...Vögtlin Instruction Manual SmartTrak Series 100 107 private static uint CalcCRC byte cmnd ...
Страница 111: ...Vögtlin Instruction Manual SmartTrak Series 100 109 Else ...
Страница 112: ...Vögtlin Instruction Manual SmartTrak Series 100 110 crcWord crcWord 2 ...