NOTE: Pins 1, 3, and 10 are connected
together inside the instrument. Do not tie these
grounds together outside the instrument. Must
have one connection per analog ground.
Recommended use listed.
Instrument Power:
The SmartTrak requires a 15-24 VDC power supply. If you are using the power supply
supplied by Vögtlin, connect it to the 15-pin HD DB-15 connector on the side of the
instrument. If you are supplying your own power source, it must be a regulated 15-24 VDC
with ripple not to exceed 100 mV peak-to-peak. It must be capable of producing the current
specified for the appropriate voltage shown in Figure 2-5: Power Supply Requirements.
Apply power as follows: positive (+) to the green (pin 6) and negative (-) to the yellow (pin
5) wires. The instrument is polarity sensitive. If you reverse this wiring, the instrument will
not be damaged, but it will not function.
Instrument Grounding:
The SmartTrak has very high levels of RFI and EMI shielding built into the metal electronics
cover (meets or exceeds the CE Standard EN 61326-1; 2006). To maintain the integrity of
this CE rating, it is critical that a path be provided for any residual internal noise to exit the
instrument or it may register on the outputs. Grounding provides this path.
To properly ground your instrument, secure the chassis to solid earth ground using the
mounting holes on the bottom of the flow body. If the instrument will be used without
permanent mounting (on a laboratory bench, for instance) then, using the provided cable,
connect the shield wire (no insulation) to earth ground in your facility. If you purchased a
Vögtlin power supply, a ground wire is provided for your convenience.
Figure 2-5: Power Supply Requirements
Instrument Type
Recommended Input
Voltage
Minimum Current
Required (mA)
M100L Meter
15-24 VDC (+ 10%)
230
M100M Meter
15-24 VDC (+ 10%)
230
M100H Meter
15-24 VDC (+ 10%)
230
C100L Controller
24 + 10% VDC
500
C100M Controller
24 + 10% VDC
800
C100H Controller
24 + 10% VDC
1260
NOTE:
The Compod™ adds an additional 100mA to the ST2 current ratings
.
Содержание SmartTrak 100 Series
Страница 63: ...63 C ADJUSTMENTS Clicking on adjustments brings up the meter settings screen ...
Страница 74: ...72 Appendix A C100 M100 SmartTrak Product Specifications Dimensions and Mounting ...
Страница 75: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 73 ...
Страница 76: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 4 7 ...
Страница 77: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 75 ...
Страница 78: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 76 ...
Страница 79: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 77 ...
Страница 80: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 78 ...
Страница 82: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 80 ...
Страница 83: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 81 ...
Страница 85: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 83 ...
Страница 86: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 84 ...
Страница 87: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 85 ...
Страница 89: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 87 ...
Страница 90: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 88 ...
Страница 91: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 89 ...
Страница 92: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 90 ...
Страница 109: ...Vögtlin Instruction Manual SmartTrak Series 100 107 private static uint CalcCRC byte cmnd ...
Страница 111: ...Vögtlin Instruction Manual SmartTrak Series 100 109 Else ...
Страница 112: ...Vögtlin Instruction Manual SmartTrak Series 100 110 crcWord crcWord 2 ...