Pre-Installation Check List
1.
Double-check to be sure that the O-ring material used in your instrument is
compatible with the gas to be measured.
The O-ring material used in your
SmartTrak can be found in the Data Label. See Appendix A for a table of elastomer
compatibility with a wide variety of gases.
2.
Vögtlin strongly recommends you install an in-line filter upstream of the
instrument.
Recommended filter size: 10 micron. A 10 micron filter is available
from Vögtlin as an accessory. See Appendix B or contact your local Vögtlin
distributor.
3.
Do not locate the instrument in areas subject to sudden temperature changes,
excessive moisture or near equipment radiating significant amounts of heat.
Be sure to allow adequate space for cable connectors and wiring.
4.
For controllers, use a properly sized pressure regulator.
Make sure the pressure
regulator is not too small or too big. There can be no restrictions (such as valves,
tubing or pipe internal diameters, reducers, etc.) upstream or downstream of the
controller with a dimension that is less than the valve orifice diameter. To
determine orifice diameter, consult the calibration certificate included with your
instrument. If restricted, controller will not reach full scale.
5.
Output Signals:
The SmartTrak has two analog outputs that are linearly
proportional to the gas mass flow rate, one voltage and one current. Choose from 0-
20 mA or 0/0/4-20 mA for current plus one voltage signal: 0-5 VDC or 0-10 VDC
or 1-5 VDC. The output signals specified at time of order will be indicated on the
data label. You may change among the current and the voltage output signal at
your discretion after receipt of the instrument using the Pilot Module or the
SmartTrak Software (see Chapters 4 & 5). Changing the output signals has no
influence on the instrument’s accuracy.
NOTE: For the SmarTrak 100 plus Compod, the 0-5 VDC I/O cannot be used. Only
0/4-20 mA I/O is available.
6.
The CAT-5
connector on the side of the SmartTrak is NOT an Ethernet connector.
It is for use with the optional Remote Pilot Module. Do not plug an Ethernet cable
here as damage may result. Keep this connector covered if possible whenever it is
not in use.
Содержание SmartTrak 100 Series
Страница 63: ...63 C ADJUSTMENTS Clicking on adjustments brings up the meter settings screen ...
Страница 74: ...72 Appendix A C100 M100 SmartTrak Product Specifications Dimensions and Mounting ...
Страница 75: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 73 ...
Страница 76: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 4 7 ...
Страница 77: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 75 ...
Страница 78: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 76 ...
Страница 79: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 77 ...
Страница 80: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 78 ...
Страница 82: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 80 ...
Страница 83: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 81 ...
Страница 85: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 83 ...
Страница 86: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 84 ...
Страница 87: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 85 ...
Страница 89: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 87 ...
Страница 90: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 88 ...
Страница 91: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 89 ...
Страница 92: ...Vögtlin Instruments Instruction Manual SmartTrak 100 Series 90 ...
Страница 109: ...Vögtlin Instruction Manual SmartTrak Series 100 107 private static uint CalcCRC byte cmnd ...
Страница 111: ...Vögtlin Instruction Manual SmartTrak Series 100 109 Else ...
Страница 112: ...Vögtlin Instruction Manual SmartTrak Series 100 110 crcWord crcWord 2 ...