EMC GUIDANCE

EMC GUIDANCE

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity

Immunity test

Compliance 
level

IEC 60601 
Test level

Conducted RF
IEC 61000-4-6 

Radiated RF 
IEC 61000-4-3

10V/m, 80% Am
at 1kHz

150 kHz to
80 MHz: 
3 Vrms
6Vrms (in ISM 
and amateur 
radio bands)
80% Am at 1kHz 

Electromagnetic environment - guidance

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 
strength in the location in which the device is used exceeds the applicable RF compliance level 
above, the device should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating the device.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

b

The device is intended for use in the electromagnetic environment specified below. The customer or the 
user of the device should assure that it is used in such an environment.

150 kHz to
80 MHz: 
3 Vrms
6Vrms (in ISM 
and amateur 
radio bands)
80% Am at 1kHz 

Portable and mobile RF communications 
equipment should be used no closer to any part 
of the device, including cables, than the 
recommended separation distance calculated 
from the equation appropriate for the frequency 
of the transmitter.
Recommended separation distances:
d=0.35        ;
d=1.2

10V/m, 80% Am
at 1kHz

80 MHz to 800 MHz:
d=1.2
800 MHz to 2.7 GHz:
d=2.3

where,

 P

 is the maximum 

output power rating of the 
transmitter in watts (W) 
according to the 
transmitter manufacturer, 

d

 is the recommended 

separation distance in 
meters (m).  Field 
strengths from fixed RF 
transmitters, as 
determined by an 
electromagnetic site 
survey, should be less 
than the compliance level 
in each frequency range.
Interference may occur in 
the vicinity of equipment 
marked with the following
symbol:

P

P

P

P

a

b

 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.7 GHz 

Table 4

Recommended separation distances between portable and mobile RF communications 
equipment and the device.

Rated maximum output

power of transmitter

(W) 

Separation distance according to frequency of transmitter (m)

0.01

0.1

1

10

100

0.12

0.12

0.38

1.2

3.8

12

0.23

0.73

2.3

7.3

23

=

d

=

d

1.2

2.3

=

d

0.38

1.2

3.8

12

3.5

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the device can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF communications equipment 
(transmitters) and the device as recommended below, according to the maximum output power of the 
communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance 

d

 in meters (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where 

P

 is the maximum output power rating of the transmitter in watts (W) according to the 

transmitter manufacturer.
NOTE 1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.

P

P

P

Содержание TMB-1776

Страница 1: ...Pressure Monitor TMB 1776 Please read the user s manual carefully and thoroughly so as to ensure the safe usage of this product Keep the manual well stored for further reference in case you have probl...

Страница 2: ...ds Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances Contraindications ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood p...

Страница 3: ...igns below might be in the user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for COMPLIES WITH MDD 93 42 EEC REQ...

Страница 4: ...ation When measuring please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disc...

Страница 5: ...two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact your LOCAL DEALER for SERVICE Don t open or repair the device by yourself...

Страница 6: ...he battery cover Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up The Choice of Power...

Страница 7: ...re using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory The setting range of the year 2017 2057 time format available 24H 1 When the monit...

Страница 8: ...rtably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep br...

Страница 9: ...ll finish the whole measurement Take user 1 for example 2 Press the to power off otherwise it will turn off within 1 minute The current No is No 1 The corresponding time is 6 00 24H Format The corresp...

Страница 10: ...or 3 seconds when the monitor is in the memory recall mode dEL ALL will flash on the display 2 Press to confirm deleting and dEL donE will be shown on the display then the monitor will turn off 3 If y...

Страница 11: ...he device is not suitable for use on patients with implanted electrical devices such as cardiac pacemakers and defibrillators Ask your Doctor before use 3 We suggest above mentioned population to meas...

Страница 12: ...arging Systolic relax blood entering Diastolic When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When...

Страница 13: ...ement can affect the measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the...

Страница 14: ...anual General Requirements for Safety Electromagnetic compatibility Performance requirements Clinical investigation Usability Software life cycle processes Bio compatibility ISO 10993 1 2009 Biologica...

Страница 15: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 16: ...ion appropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output po...

Страница 17: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: