6VDC 4×AA batteries

Approx.258g(Excluding the batteries and cuff)

A01

Digital LCD display  V.A.80mm × 58.5mm

Approx.140.4mm×110.4mm×64.8mm

4×AA batteries, user manual, 

flannelette bag

6V      1A (Not included)

Battery powered mode: 

AC adaptor powered mode:

(Please only use the recommended  AC 
adaptor model). 

    This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your blood  
pressure monitor. If the product is not operating as you think it 
should, check here before arranging for servicing.

PROBLEM SYMPTOM

CHECK THIS

REMEDY

No power

Low
batteries

Error
message

Display will not 
light up.

Batteries are exhausted. Replace with new batteries

Insert the batteries 
correctly

Replace with new batteries

Batteries are inserted
incorrectly.

Display is dim or 
shows

Batteries are low.

“E 01” shows

“E 03” shows

“E 04” shows

“EExx”, shows 
on the display.

A calibration error
occurred. (XX can
be some digital symbol,
such as 01, 02,etc., if 
this similar situation 
appear, all belong to 
calibration error.)

Retake the measurement.
If the problem persists,
contact the retailer or our
customer service 
department for further 
assistance.Refer to the 
warranty for contact 
information and return 
instructions.

The treatment of the
measurement failed.

AC adaptor is inserted
incorrectly.

Insert the AC adaptor 
tightly

“E 02” shows

The cuff is too tight 
or too loose.

Refasten the cuff and then
measure again.

The monitor 
detected motion 
while measuring.

Movement can affect the
measurement.Relax for a 
moment and then 
measure again.

Relax for a moment and
then measure again.

The measurement 
process does not 
detect the pulse 
signal.

Loosen the clothing on
the arm and then 
measure again. 

Warning
message

Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.

“out” shows

Out of measurement
range

TROUBLESHOOTING

SPECIFICATIONS

About 22cm~42cm

Type BF applied part

WARNING: No modification of this equipment is allowed.

Power supply

Display mode

Measurement mode

Oscillographic testing mode

Measurement range

Measurement perimeter

of the upper arm

Weight

External dimensions

Attachment

Mode of operation

Continuous operation

Degree of protection

Protection against 

ingress of water

Accuracy

Normal working condition

Storage & transportation

condition

Software Version

Pressure:
5

-40

within±3mmHg(0.4kPa) 

Pulse value:±5%

Rated cuff pressure: 
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure: 
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) 
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute

IP21 It means the device could protected against 
solid foreign objects of 12.5mm and greater, and 
protect against vertically falling water drops.

Device Classification

Battery Powered Mode: 
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment

A temperature range of :+5°C to +40°C 
A relative humidity range of 15% to 90%, 
non-condensing, but not requiring a water vapour 
partial pressure greater than 50 hPa
An atmospheric pressure range of : 
700 hPa to 1060 hPa

Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing, 
at a water vapour pressure up to 50hPa

Содержание TMB-1776

Страница 1: ...Pressure Monitor TMB 1776 Please read the user s manual carefully and thoroughly so as to ensure the safe usage of this product Keep the manual well stored for further reference in case you have probl...

Страница 2: ...ds Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances Contraindications ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood p...

Страница 3: ...igns below might be in the user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for COMPLIES WITH MDD 93 42 EEC REQ...

Страница 4: ...ation When measuring please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disc...

Страница 5: ...two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact your LOCAL DEALER for SERVICE Don t open or repair the device by yourself...

Страница 6: ...he battery cover Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up The Choice of Power...

Страница 7: ...re using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory The setting range of the year 2017 2057 time format available 24H 1 When the monit...

Страница 8: ...rtably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep br...

Страница 9: ...ll finish the whole measurement Take user 1 for example 2 Press the to power off otherwise it will turn off within 1 minute The current No is No 1 The corresponding time is 6 00 24H Format The corresp...

Страница 10: ...or 3 seconds when the monitor is in the memory recall mode dEL ALL will flash on the display 2 Press to confirm deleting and dEL donE will be shown on the display then the monitor will turn off 3 If y...

Страница 11: ...he device is not suitable for use on patients with implanted electrical devices such as cardiac pacemakers and defibrillators Ask your Doctor before use 3 We suggest above mentioned population to meas...

Страница 12: ...arging Systolic relax blood entering Diastolic When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When...

Страница 13: ...ement can affect the measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the...

Страница 14: ...anual General Requirements for Safety Electromagnetic compatibility Performance requirements Clinical investigation Usability Software life cycle processes Bio compatibility ISO 10993 1 2009 Biologica...

Страница 15: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 16: ...ion appropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output po...

Страница 17: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: