VMAX TMB-1776 Скачать руководство пользователя страница 5

LCD display signal

SYMBOL

DESCRIPTION

Systolic pressure

High blood pressure 

Diastolic pressure

Low blood pressure

EXPLANATION

mmHg

Measurement Unit of the blood pressure

Low battery

Batteries are low and need to be replaced

Current Time

Year/Month/Day, Hour : Minute

(1mmHg=0.133kPa)

Heartbeat

Irregular heartbeat 

Average value

The average value of blood pressure

Pulse in beats per minute

Pulse display

Blood pressure monitor is detecting an 
irregular heartbeat during measurement.

Blood pressure 
level indicator

Indicate the blood pressure level

Blood pressure monitor is detecting a 
heartbeat during measurement.

/min

Memory

Indicate it is in the memory mode and 
which group of memory it is.

7

6

CAUTION

* It is recommended that the performance should be checked every 2 years and after 
maintenance and repair, by retesting at least the requirements in limits of the error of the 
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT 
according to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists, 
descriptions, calibration instructions, etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulate the device from the main supply. Do not position the 
device in a position where it is difficult to disconnect from the supply mains to safely 
terminate operation of ME equipment.
* The operator shall not touch output of batteries /adapter and the patient simultaneously.
* Cleaning: Dust environment may affect the performance of the unit. Please use the soft 
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 
cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using, 
please contact your LOCAL DEALER for SERVICE. Don’t open or repair the device by 
yourself in the event of malfunctions. The device must only be serviced, repaired and 
opened by individuals at authorized sales/service centers.
* Please report to your LOCAL DEALER if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage 
temperature between uses until it is ready for intended use. At least 30 min required for ME 
equipment to cool from the maximum storage temperature between uses until it is ready for 
intended use.
* This equipment needs to be installed and put into service in accordance with the 
information provided in the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile 
phones, cordless telephones and their base stations, walkie-talkies can affect this 
equipment and should be kept at distance “d” from the equipment. The “d” is calculated by 
the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 
60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable parts specified/ authorized by 
MANUFACTURER. Otherwise, it may cause damage to the unit or danger to the 
user/patients.
* There is no “luer lock” connectors used in the construction of tubing, there is a possibility 
that they might be inadvertently connected to intravascular fluid systems, allowing air to be 
pumped into a blood vessel.
* Please use the device under the environment which Is provided in the user manual. 
Otherwise, the performance and lifetime of the device will be impacted and reduced.

INTRODUCTION

INTRODUCTION

Содержание TMB-1776

Страница 1: ...Pressure Monitor TMB 1776 Please read the user s manual carefully and thoroughly so as to ensure the safe usage of this product Keep the manual well stored for further reference in case you have probl...

Страница 2: ...ds Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances Contraindications ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood p...

Страница 3: ...igns below might be in the user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for COMPLIES WITH MDD 93 42 EEC REQ...

Страница 4: ...ation When measuring please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disc...

Страница 5: ...two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact your LOCAL DEALER for SERVICE Don t open or repair the device by yourself...

Страница 6: ...he battery cover Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up The Choice of Power...

Страница 7: ...re using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory The setting range of the year 2017 2057 time format available 24H 1 When the monit...

Страница 8: ...rtably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep br...

Страница 9: ...ll finish the whole measurement Take user 1 for example 2 Press the to power off otherwise it will turn off within 1 minute The current No is No 1 The corresponding time is 6 00 24H Format The corresp...

Страница 10: ...or 3 seconds when the monitor is in the memory recall mode dEL ALL will flash on the display 2 Press to confirm deleting and dEL donE will be shown on the display then the monitor will turn off 3 If y...

Страница 11: ...he device is not suitable for use on patients with implanted electrical devices such as cardiac pacemakers and defibrillators Ask your Doctor before use 3 We suggest above mentioned population to meas...

Страница 12: ...arging Systolic relax blood entering Diastolic When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When...

Страница 13: ...ement can affect the measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the...

Страница 14: ...anual General Requirements for Safety Electromagnetic compatibility Performance requirements Clinical investigation Usability Software life cycle processes Bio compatibility ISO 10993 1 2009 Biologica...

Страница 15: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 16: ...ion appropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output po...

Страница 17: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: