background image

24

25

Complied Standards List

EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Application of 

risk management to medical devices

EN 1041:2008 Information supplied by the manufacturer of medical 

devices

EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical 

electrical equipment - Part 1: General requirements for basic safety and 
essential performance

EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical 

equipment - Part 1-11: General requirements for basic safety and 
essential performance - Collateral standard: Requirements for medical 
electrical equipment and medical electrical systems used in the home 
healthcare environment

EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical equipment 

- Part 1-2: General requirements for basic safety and essential 
performance - Collateral standard: Electromagnetic disturbances - 
Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: 

Requirements and test methods for non-automated measurement type

EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - Part 3: 

Supplementary requirements for electro-mechanical blood pressure 
measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test 

procedures to determine the overall system accuracy of automated 
non-invasive sphygmomanometers

EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 Medical 

electrical equipment - Part 1-6: General requirements for basic safety 
and essential performance - Collateral standard: Usability

 IEC 62366-1:2015 Medical devices - Part 1: Application of usability 

engineering to medical devices

EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015
Medical device software - Software life-cycle processes

Risk management

Labeling

User manual

General Requirements 
for Safety

Electromagnetic
compatibility

Performance
requirements

Clinical investigation

Usability

Software life-cycle 
processes

Bio-compatibility

ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: 

Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: 

Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices - Part 

10: Tests for irritation and skin sensitization

EN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. Symbols 

to be used with medical device labels, labelling and information to be 
supplied. Part 1 : General requirements

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 

2-30: Particular requirements for the basic safety and essential 
performance of automated non-invasive sphygmomanometers

ISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: 

Clinical validation of automated measurement type

Contact Information

For more information about our products, please visit www.vie-max.com. You can 

get customer service, usual problems and customer download, Vie-max will serve you 
anytime.

Authorized European Representative:

Manufactured by:

Company:

Address:

Company:

MDSS - Medical Device Safety Service GmbH

Address:

Schiffgraben 41, 30175 Hannover, Germany

International Distributor : VIE MAX scs
www.vie-max.com
[email protected]
 
Distributor in Lebanon : HOSPISERVICES / Beirut
[email protected]
Phone : +961-9-237 987
Hotline: +961-9-666 909

Authorized Component

1

.

Please use the

authorized adapter (Optioanl, Not included).

Guangdong Transtek Medical Electronics Co., Ltd.

Zone A, No.105 ,Dongli Road, Torch Development District, 
Zhongshan,528437,Guangdong, P.R.C.

Guangdong Transtek Medical Electronics Co., Ltd.

Adapter
Model

BLJ06L060100P-B

Input

AC 100-240V

50/60Hz  0.2A Max

Output

6V      1000mA

AUTHORIZED COMPONENT

COMPLIED STANDARDS LIST

Содержание TMB-1776

Страница 1: ...Pressure Monitor TMB 1776 Please read the user s manual carefully and thoroughly so as to ensure the safe usage of this product Keep the manual well stored for further reference in case you have probl...

Страница 2: ...ds Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances Contraindications ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood p...

Страница 3: ...igns below might be in the user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for COMPLIES WITH MDD 93 42 EEC REQ...

Страница 4: ...ation When measuring please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disc...

Страница 5: ...two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact your LOCAL DEALER for SERVICE Don t open or repair the device by yourself...

Страница 6: ...he battery cover Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up The Choice of Power...

Страница 7: ...re using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory The setting range of the year 2017 2057 time format available 24H 1 When the monit...

Страница 8: ...rtably with your tested arm resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep br...

Страница 9: ...ll finish the whole measurement Take user 1 for example 2 Press the to power off otherwise it will turn off within 1 minute The current No is No 1 The corresponding time is 6 00 24H Format The corresp...

Страница 10: ...or 3 seconds when the monitor is in the memory recall mode dEL ALL will flash on the display 2 Press to confirm deleting and dEL donE will be shown on the display then the monitor will turn off 3 If y...

Страница 11: ...he device is not suitable for use on patients with implanted electrical devices such as cardiac pacemakers and defibrillators Ask your Doctor before use 3 We suggest above mentioned population to meas...

Страница 12: ...arging Systolic relax blood entering Diastolic When ventricles contract and pump blood out of the heart the blood pressure reaches its maximum value in the cycle which is called systolic pressure When...

Страница 13: ...ement can affect the measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the...

Страница 14: ...anual General Requirements for Safety Electromagnetic compatibility Performance requirements Clinical investigation Usability Software life cycle processes Bio compatibility ISO 10993 1 2009 Biologica...

Страница 15: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 16: ...ion appropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output po...

Страница 17: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: