09060 Issue 1
33
Page
of 40
EN
Alpha - Instructions for Use
EN 60601-1-2:2015 - Immunity tests
Immunity test
Test level
Compliance level
Reached
Electrostatic
discharge (ESD)
EN 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Electrical fast
transient/burst EN
61000-4-4
±2kV for power supply lines
+/- 2kV
Surge
EN 61000-4-5
± 0.5 kV, ±1kV differential mode
Device is identified as Class II
unearthed equipment therefore
common mode testing is omitted.
+/- 0.5 kV, 1kV
Voltage dips, short
interruptions on
power supply input
lines EN 61000-
4-11
100% drop, 0.5 cycles, 0
o
, 45
o
, 90
o
,
135
o
, 180
o
, 225
o
, 270
o
, 315
o
100% dip, 1 cycle
30% dips, 25/30 cycles
240V AC, 50Hz
100V AC, 50Hz
100% drop, 0.5
cycles, 0
o
, 45
o
,
90
o
, 135
o
, 180
o
,
225
o
, 270
o
, 315
o
100% dip, 1 cycle
30% dip, 25/30
cycles
240V AC, 50Hz
100V AC, 50Hz
Conducted RF
EN 61000-4-6
3V/m 0.15-80MHz
6V/m ISM/amateur radios bands
0.15-80MHz
80% AM at 1kHz
3V/m 0.15-80MHz
6V/m ISM/
amateur radios
bands
0.15-80MHz
80% AM at 1kHz
Radiated RF
EN 61000-4-3
3V/m (professional healthcare)
80MHz to 2700MHz
3V/m
(professional
healthcare)
80MHz to
2700MHz
Proximity fields
from RF devices
EN 61000-4-3
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
Power frequency
(120V, 60Hz and
240V, 50Hz)
magnetic field
30A/m
30A/m