09060 Issue 1
15
Page
of 40
EN
Alpha - Instructions for Use
Non-reproducible syringe strokes (red syringe, after
the 5
th
Stroke). Repeat calibration verification.
Outside 3%, 6% or 25% (purple syringe, and shows
>3%, >6% or >25% in top right corner of screen).
The % it is out by indicates the severity of the issue.
Over 3%, repeat calibration verification.
Over 6% may indicate the device requires cleaning or
maintenance.
Over 25%, the user should contact technical support.
The following displays after the calibration update:
Calibration update (purple syringe with purple tick
in a circle, % shown in top right corner of screen).
A report may be printed or saved for the calibration verification if required, see
section 3.3. Reporting.
If the procedure was followed correctly and the error icon is showing, the
calibration verification may be repeated. If the error continues to show, contact
the manufacturer using contact information at the start of this document.
3.5. Connection to Electronic Medical Records (EMR) via Vitalograph
Connect
Connection to an EMR system may be achieved using the Connect
service. This is available on the Flash Drive PN 93002 supplied with the
device. Refer to the Connect IFU PN 09570 for information on installing
and running Connect.
3.5.1.
Connect Configuration
On the main menu select ‘Connect Config’
.
The following information is required:
-
Address: URL for the Connect service - http://{host}[:{port}]/
connect
-
Network information: Three options - disabled, wired or wireless.