3.4.
Calibration Verification
ISO 26782 recommendations require that the difference between
the volume measured by the spirometer and the volume pumped
into the spirometer from a syringe is within 3% (2.5% for the device with
an additional 0.5% allowed for the syringe).
1.
Attach the flowhead to the syringe as per figure 6.
Figure 6
2.
Select ‘Configuration’ on the main menu, select ‘Calibration Verification’
option.
3. Enter syringe details: serial number, syringe volume (will default to 3L),
pressure, altitude and humidity %. At a minimum the serial number,
syringe volume and temperature must be entered. The temperature
auto fills from the internal temperature sensor but may be changed if
required. If a calibration verification was performed previously this will
pre-populate with the details of last syringe used.
Press Continue.
4. Complete 3 full strokes as per onscreen instructions.
5. If the reproducibility between the 3 consecutive syringe strokes is less
than 3% of 3L, the calibration verification has passed, and a green
tick displays on the syringe icon and <3% is displayed in the top right
corner.
In the unlikely event the result is outside 3% or non-reproducible, an
error icon will display.
Successful verification check with tick in syringe
indicating a pass.