Do NOT use AST if:
- You think your blood glucose is low.
- You are unaware of hypoglycemia.
- You are testing for hyperglycemia.
- Your AST results do not match the way you feel.
- Your routine glucose results often fluctuate.
Preparing the Puncture Site
Stimulating blood flow by rubbing the puncture site before
blood extraction has a significant influence on the glucose value
obtained.
Blood from a site that has not been rubbed exhibits a measurably
different glucose concentration than blood from a site that has.
When the puncture site was rubbed prior to blood extraction, the
difference was significantly reduced.
Reference : Jungheim, Karsten, and Theodor Koschinsky. “Glucose Monitoring at the Arm Risky de-
lays of hypoglycemia and hyperglycemia detection.” Diabetes Care 25.6 (2002): 956-960.
Please follow the suggestions below before obtaining a drop
of blood:
- Wash and dry your hands before starting.
- Select the puncture site at fingertips or another body part.
- Rub the puncture site for about 20 seconds before penetration.
- Clean the puncture site using cotton moistened with 70% alcohol
and let it air dry.
- Replace the lancing device cap with the clear cap for AST. Pull
the cocking control back until it clicks. When lancing the forearm
or hand, avoid lancing the areas with obvious veins because of
excessive bleeding.
4.6. DOWNLOAD APPLICATION
BW-DGM1 is designed to be used in conjunction with the
BewellConnect® application, available in the App Store or Google
Play for free download.
The following information is subject to
change.
bw
My
Download the BewellConnect® app on your smartphone or
tablet : App store or Google Play.
Then click on the MyGluco icon.
It is simple and intuitive to use, for better understanding of your
current condition and to achieve better diabetes control.
System Requirement : iOS (8 or above) / Android (4.3 or above)
5. HOW TO MAINTAIN AND STORE THE UNIT
5.1 REPLACING THE BATTERY
The battery compartment is located on the back of the device.
1. Open the battery compartment
2. Replace the dead battery, taking care to observe the correct
polarity. Incorrect polarity can cause damage and compromise
the warranty of your appliance.
3. Close the battery compartment. Make sure the battery is inserted