The following are examples of situations where the Mini Console would be useful.
Example 1
The desk/counter at control room technician’s workstation is crowded making it difficult to
find room for the 94A-07 console. In this case, the 94A-07 console could be located where
more space is available, possibly at the end of the workstation or on a shelf, and the smaller
Mini Console could be placed in front of the technician or even beneath the counter at the
technician’s location.
Example 2
There is the need for a secondary location in the control room for personnel to have two-way
communication with the procedure room and this location is too far from the 94A-07 console
for it to be practical to connect a second headset to the 94A-07 console directly. This might be
across the room or in a separate room or office. In this case, the Mini Console would duplicate
the 94A-07 console’s intercom functionality. An additional headset and foot switch and/or
desk microphone will be needed. If located in a separate room from the 94A-07 console, and
a desk microphone is to be used with the Mini Console, an additional communication speaker
will also need to be installed in the room with the Mini Console. An additional communication
speaker is not necessary if a headset is used. Please contact Vis-A-Vis at 800-319-6014 or by
email at [email protected] to discuss your specific requirements.
7.2.2 Installation
1. Determine where the Mini Console will be located. Possibilities include but are not limited
to on top or underneath a desk surface (see images below). If it is to be located under a
countertop, it can be secured using two 2 1/4 inch #8 wood screws through pre-existing
holes in the sheet metal enclosure. Before installation, make sure this screw length is not
too long. If too long, it may penetrate through the countertop and damage the finish.
2. Run an Ethernet cable between the jack labeled
BLACK
on the rear of the 94A-07 console
and the jack labeled
BLACK IN
on the rear of the Mini Console. This is the only cable
that needs to be run between the two consoles. In addition to communication, this cable
provides power from the 94A-07 console to the Mini Console. A 100’ cable is provided.
VITALINQ 94A
CONSOLE
VITALINQ MC-11 console shown mounted
below control room workstation countertop
Control room workstation countertop
VITALINQ 94A
CONSOLE
VITALINQ MC-11 console shown mounted
above control room workstation countertop
Control room workstation countertop
VITALINQ 94A
CONSOLE
VITALINQ MC-11 console shown mounted
below control room workstation countertop
Control room workstation countertop
VITALINQ 94A
CONSOLE
VITALINQ MC-11 console shown mounted
above control room workstation countertop
Control room workstation countertop