Vpod Instruction Manual 

 

wrapping  the  sensor  too  tight.  Inspect  the  sensor  site  to  ensure  the  skin’s  integrity  and  the 

adhesion position of the sensor is correct. More frequent inspection should be taken if necessary. 

Loss of pulse signal can occur in any of the following situations: 

a)  The sensor is too tight; 
b)  There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or 

sunlight; 

c)  A blood pressure cuff is inflated on the same extremity as the one to which an SpO

2

 sensor is 

attached. 

4.4 Display mode 

The handheld pulse oximeter uses an OLED display for a readout. It can display the SpO

2

 and pulse 

rate (PR) value, as well as a pulse column and SpO

2

 waveform.   

There  are  three  display  modes  shown  in  Fig  7.  The  first  figure  is  pulse  column  display  mode.  The 

second figure is the waveform mode. The third figure is line waveform mode indicating SpO

2

% trend.   

     

     

   

                      (1)                                        (2)                                          (3) 

Fig.7 Three display modes 

 

SpO

2

:

    Percent oxygen saturation value displayed above is 98%   

 

PR:

    Pulse rate value displayed is 72 bpm   

  Fig.7(1) 

Pulse column:

 This is used for signal identification and quality indication during motion 

and  low  signal  to  noise  situations.  The  bar  rises  and  falls  with  the  pulse,  its  height  indicating 

signal quality. When the bar is very low, the SpO

2

 and pulse rate values may be suspect. 

  Fig.7(2) 

Filling-up  SpO

2

  plethysmogram  wave:

  SpO

2

  plethysmogram  measurement  is 

employed to determine the oxygen saturation of hemoglobin in the arterial blood. If , for example, 

98% hemoglobin molecules in the red blood cells of the arterial blood combine with oxygen, then 

the blood has a SpO

2

 oxygen saturation of 98%. The SpO

2

 numeric on the Oximeter will read 

98%, refer to Fig.7.(2). The SpO

2

 numeric shows the percentage of hemoglobin molecules which 

have  combined  with  oxygen  molecules  to  form  oxyhemoglobin.  The  SpO

2

  plethysmogram 

measurement can also provide a plethysmogram wave. During this mode, the PR tone is dumb.   

  Fig.7(3)   

SpO

2

 plethysmogram wave

 

 

PR  tone  modulation

:  Beeps  in  sync  with  the  patient’s  pulse,  even  under  most  challenging 

patient motion conditions. 

CHAPTER 5 Other information 
5.1 Alarm 
Alarm:

 Technical alarm and physiological alarm. 

Technical alarm:

 finger out, probe off, power low and error code. 

In the situation that the finger is not inserted correctly or the connection state of the probe is not good 

results in failure of measurement, “Finger out” or ”Sensor off” may be displayed on the normal screen. 

When battery power is lower than 2.7

±

0.1V, the sign 

 

  will flicker in its display area. Replace the 

batteries as soon as possible. 
In the failure state, the oximeter will display error codes, and will automatically power off if the error 

code  display  lasts for more  than 8  seconds.  For  the details  and  definitions  on  error,  please  refer  to 

chapter 4.2. 

Physiological alarm: 

SpO

2

 and PR 

If  the  measured  SpO

2

  and/or  PR  value  is  beyond  the  default  limit,  alarm  will  be  activated,  and  the 

corresponding value will flash with audio alarm “didi—didi….” . 

By default:   

SpO

2

: The upper limit: 100% 

The lower limit: 90%   

PR:    The upper limit: 100bpm 

The lower limit: 60bpm 

Note:  during  the  alarm  is  issued,  long press the functional  button, you  can  silence  the  alarm for 30 

seconds.   

5.2 About the button 

1) Definition of button 

There are three buttons in the oximeter: Power button, Function button and Setting button. 

Power button:

 This acts as a Power On switch when the unit is in an Off condition. 

Function Button:

 When the unit is On, it acts as a Function button. 

Setting button:

 Located at the right of the oximeter, it has no function when the power is off. When 

the unit is On, it acts as a Setting button. 

2) Definition of Button Press 

There are three ways to press the button: 

Short Press:

 press the button quickly, the duration time should no more than 1 second. 

Double press:

 Two-time continuous press, the time between the two press actions should be no more 

than 0.5 second. 

Содержание VPOD

Страница 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Страница 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Страница 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Страница 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Страница 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Страница 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Страница 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Страница 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Страница 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Страница 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Страница 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Страница 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...