Vpod Instruction Manual 

 

12 

APPENDIX A    Specifications 
Notices

:   

 

Specifications may be changed without prior notice. 

 

The  circuit  diagrams,  the  list  of  components,  the  illustration  of  diagrams,  and  the 

detailed  rules  of  calibration,  are  provided  exclusively  to  professional  personnel 

authorized by our company. 

 

SpO

2

 

Display Range:                0-100%   

Measurement Range:    70%-100%                                                                                     

Resolution:            1%                                                                                                 

Accuracy:              80%-100%: ±2%   

                                70%-79%: ±3% 

                                0%-69%: unspecified 

Heart (Pulse) Rate                 

Display Range:                0-254 bpm   

                                                                         

Measurement Range:    30-235 bpm   

Resolution:            1bpm                                                                                             

Accuracy:              ±2bpm or ±2% (the larger is applied) 

Alarm 

Alarm:              Technical alarm and physiological alarm   

Modes:            Visual and audible 

SpO

2

:      The upper limit: 100% 

The lower limit: 60%   

PR:        The upper limit: 100bpm 

The lower limit: 60bpm   

Display                     

Type:                OLED, double color           

Parameters:    SpO

2

, PR, SpO

2

 plethysmogram waveform, PR bar 

Mode:                3 display modes. 

Record 

Patient ID              1-10 

Data record          Up to 72 hours 

 

Data transmission 

Transmission method:    Cable Transmission   
Data Cable Interface:      DB9 (Connected to Pulse Oximeter)     
                                            USB (Connected to PC) 

Operation Environment 

Operating temperature:      5

℃~

40 

 

Relative humidity:             

80%, no condensation 

Atmosphere pressure:      86 kPa

106 kPa 

Power supply:                    Two AA alkaline batteries   

Operation time                  About 50 hours of typical operation 

Transport and Storage Environment 

Storage Temperature:      -20

55

 

Relative humidity:             

93%, no condensation 

Atmosphere pressure:      50 kPa

106 kPa 

 

Classification according to IEC60601-1 

Type of protection: Internally powered equipment 

Degree of protection: Type B-Applied Part 

Mode of Operation: Continuous 

Safety:                    IEC Standard 60601-1 

Dimensions:     

139mmX44mmX24mm (LXWXH) 

Weight:     

          110g (with alkaline batteries) 

 

 

 

 

 

 

 

 

 

 

APPENDIX B   

 

DB9 

 

USB 

       

Содержание VPOD

Страница 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Страница 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Страница 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Страница 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Страница 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Страница 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Страница 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Страница 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Страница 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Страница 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Страница 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Страница 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...

Страница 13: ...Display Type OLED double color Parameters SpO2 PR SpO2 plethysmogram waveform PR bar Mode 3 display modes Record Patient ID 1 10 Data record Up to 72 hours Data transmission Transmission method Cable...

Страница 14: ...eclaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting 61000 4 6 Radiated RF IEC 61000 4 3 3V m 80Hz to 2 5 GHz d P V1 5 3 80MHz to 800MHz d P E1 5 3 800MHz to 2 5G...

Страница 15: ...minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications eq...

Страница 16: ...Vpod Instruction Manual 15...

Отзывы: