Vpod Instruction Manual 

 

 

Manufacturer’s information. 

 

 

Low power indicator 

IPX1

 

Protected  against  dripping 

water 

1.4 Electromagnetism interference 

This  oximeter  is  designed  and  tested  in  compliance  with  the  EMC  standard,  complying  with  the 

international  standard  for  the  EMC  of  the  electronic  medical  device  -  IEC  60601-1-2.  However, 

because of the proliferation of radio frequency transmitting equipment and other sources of electrical 

noise  in  the  health-care  and  home  environments  (e.g.  cellular  phones,  mobile  two-way  radios, 

electrical  appliances)  it  is  possible  that  high  levels  of  such  interference  due  to  close  proximity  or 

strength of a source, may result in disruption of performance of this device. 
This  apparatus  complies  with  the  IEC  60601-1-2  international  standard.  The  requirements  of  this 

international standard are: CISPR11, GROP1, and CLASS B. 

1.5 Equipment classification 

Classification according to IEC-60601

 

According to the type of protection against 

Electrical shock: 

Internal electrical power source equipment 

According  to  the  degree  of  protection 

against Electrical shock: 

Type B equipment 

According  to  the  degree  of  protection 

against harmful ingress of water. 

Ordinary  equipment(enclosed  equipment  without 

protection against ingress of water) 

According  to  the  methods  of  sterilization 

or disinfection   

Non-sterilizable:  Use  of  Liquid  surface  disinfectants 

only. 

According to the mode of operation: 

Continuous operation 

Equipment  not  suitable  for  use  in  the  presence  of  a  flammable  anesthetic  mixture  air  or  with 

oxygen or nitrous oxide. 

1.6 Intended Use 

The  purpose  and  function  of  the  handheld  pulse  oximeter  is  to  indicate  measure  and  display  the 

functional  oxygen  saturation  of  arterial  hemoglobin  (SpO

2

)  and  pulse  rate  for  adults  and  pediatric 

patients  in  hospital,  ambulatory,  home,  and  EMS  (emergency  medical  service)  environments.  The 

pulse oximeter is intended for spot-checking these levels and not indicated for continuous monitoring. 

It can assist the clinician diagnostically by quickly displaying the patient’s SpO

2

% and pulse rate and 

can additionally store 72 hours of data. 

1.7 Product features 

 

Compact, lightweight design and simple operation 

 

Color OLED with adjustable backlight displays SpO

2

, pulse rate and pulse bar   

 

Up to 10 patient ID and 72-hour record storage, visual & audio alarm, low battery alarm   

 

Data transfer to PC for storage or printing 

 

Convenient 2 AA size alkaline batteries   

 

Suitable for adult and pediatric patient   

1.8 Accessory 
Standard accessories: 

Operator’s manual 

VPROBE V_SOFT Finger sensor: compatible with The VPOD 

Two AA-Size Alkaline batteries 

Subscription to Virtuox 

Optional accessories: 

1.  Finger sensor for adult   
2.  SpO

2

 sensor for pediatric and infant   

3.  Software license 
4.  Data cable for transmission to Virtuox 
5.  2 year extended warranty 
Confirm that the items listed are packed with the pulse oximeter. If any item on this list is missing or 

damaged, contact your distributor. Contact the carrier immediately if the shipping carton is damaged.   
 

Содержание VPOD

Страница 1: ...s of data storage High Resolution Pulse Oximetry Records at 1 second intervals Three display modes Special design for sleep monitoring LCD Displays SpO2 SpO2 waveform Pulse Rate Low power consumption...

Страница 2: ...ights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements of patents or other rig...

Страница 3: ...nts Basis of Operation 1 Red and Infrared ray Emitter Diode 2 Red and Infrared ray Receptor Diode 1 2 Safety Information Conception of Warning Precaution and Notice Warning Indicates a potential hazar...

Страница 4: ...ry to use the SpO2 and NIBP measurement on the same arm at the same time This could potentially affect measurement accuracy Notices Operation of this device in an electromagnetic field may influence i...

Страница 5: ...of operation Continuous operation Equipment not suitable for use in the presence of a flammable anesthetic mixture air or with oxygen or nitrous oxide 1 6 Intended Use The purpose and function of the...

Страница 6: ...utton Note If no press on any button for 30 seconds the oximiter will power off automatically 4 The Setting button Press this button repeatedly to increase the parameter under adjustment by one decrem...

Страница 7: ...as still as possible 3 2 Enter Setting mode Press the function button for about 4 second extended press the oximeter will enter into Setting mode and you will find a parameter item which include its t...

Страница 8: ...h After turning on the oximeter each time the function button is pressed the oximeter will switch to another display mode shown as Fig 7 4 3 Factors that may affect the measurement During operation th...

Страница 9: ...gram measurement can also provide a plethysmogram wave During this mode the PR tone is dumb Fig 7 3 SpO2 plethysmogram wave PR tone modulation Beeps in sync with the patient s pulse even under most ch...

Страница 10: ...disinfecting 1 Clean or disinfect the sensor before each use 2 Unplug the sensor from the oximeter before cleaning and or disinfecting 3 Clean the external sensor surfaces and patient contact surface...

Страница 11: ...ted 2 The connection between the Probe and Oximeter is loose 1 Connect the sensor 2 Please check if the probe was connected with oximeter correctly 5 5 1 Service Method a Service hours 9 00am 5 30pm M...

Страница 12: ...he user s request Return Policy Warranty and non warranty returns should be handled in the following manner Contact the Technical Support Department and obtain a RMA Return Materials Authorization num...

Страница 13: ...Display Type OLED double color Parameters SpO2 PR SpO2 plethysmogram waveform PR bar Mode 3 display modes Record Patient ID 1 10 Data record Up to 72 hours Data transmission Transmission method Cable...

Страница 14: ...eclaration Electromagnetic Immunity for Equipment and Systems that are not Life Supporting 61000 4 6 Radiated RF IEC 61000 4 3 3V m 80Hz to 2 5 GHz d P V1 5 3 80MHz to 800MHz d P E1 5 3 800MHz to 2 5G...

Страница 15: ...minimum distance between portable and mobile RF communication equipment transmitters and the Handheld Pulse Oximeter as recommended below according to the maximum output power of the communications eq...

Страница 16: ...Vpod Instruction Manual 15...

Отзывы: