Vpod Instruction Manual
14
situation
for
radio
(cellular/cordless)
telephones
and land mobile radios, amateur
radio,
AM
and
FM
radio
broadcast and TV broadcast
cannot be predicted theoretically
with accuracy. To assess the
electromagnetic
environment
due to fixed RF transmitters, an
electromagnetic
site
survey
should be considered. If the
measured field strength in the
location in which the Handheld
Pulse Oximeter is used exceeds
the applicable RF compliance
level above, the Handheld Pulse
Oximeter should be observed.
Recommended separation distances
between portable and mobile RF
communications equipment and the
Handheld Pulse Oximeter
The Handheld Pulse Oximeter is
intended for use in an electromagnetic
environment in which radiated RF
disturbance
are
controlled.
The
customer or the user of the Handheld
Pulse Oximeter can help prevent
electromagnetic
interference
by
maintaining
a
minimum
distance
between portable and mobile RF
communication
equipment
( transmitters) and the Handheld Pulse
Oximeter as recommended below,
according to the maximum output
power
of
the
communications
equipment.
Separation distance
according to frequency of
transmitter (m)
Rated
maxim
um
output
power
of
transmi
tter
(W)
150KHz
to
80
MHz
d=
P
V
1
5
.
3
80MHz
to 800
MHz
d=
P
E
1
5
.
3
800MH
z to 2.5
GHz
d=
P
E
1
7
0.01
0.1167
0.1167
0.2334
0.1
0.3689
0.3689
0.7378
1
1.1667
1.1667
2.3334
10
3.6893
3.6893
7.7386
100
11.666
7
11.6667
23.3334
For transmitters rated at a maximum
output power not listed above, the
recommended separation distance in
meters (m) can be estimated using the
equation applicable to the frequency of
the transmitter, where P is the
maximum output power rating of the
transmitter in watts (W) according to
the transmitter manufacturer.
NOTE1
At 80MHz and 800MHz, the
separation distance for the higher
frequency range applies.
NOTE2
These guidelines may not
apply in all situations. Electromagnetic
interference is affected by absorption
and reflection from structures, objects
and people.