36
4.2 Main Unit Cleaning
4.3 Temperature Probe Cleaning and Disinfection
4.1 Filter Replacement
4. Product Maintenance
The device should be properly maintained throughout its expected service life. Visual
inspection should be conducted on heater plate, temperature probe, power cord and
all accessories in use, on a regular monthly basis, by trained technical personnel
familiar with the technical performance of the inspired™ O2FLO. If necessary, damaged
parts should be replaced. Check that the filter is clean before using the device.
Warning: Ensure that the device is powered off and disconnected from power
supply before any cleaning or maintenance.
Warning: When the device is not in use, please keep the inlets clean, and store the
device in its original packaging.
The air inlet has a filter which needs to be replaced at least once a month. Once
discoloration appears, we would recommend the filter to be changed immediately.
The filter should be replaced every 30 days for single-patient use, or after every patient.
The replacement steps are as follows:
a. Rotate counter-clockwise to open the filter cover.
b. Remove the filter.
c. Clean the filter cover with fresh water.
d. Dry the filter cover and filter enclosure with a clean, dry cloth.
e. Replace the old filter with a new filter.
f. Rotate clockwise to close the filter cover.
* Before installing the filter, ensure that the filter is in good condition without defects.
Clean main unit, heater connection cable and VUN-300 connection cable with 75%
ethanol solvent to wipe the surface of equipment.
Warning: Do not immerse the main unit in liquid.
Warning: Do not use corrosive solvents to clean the surface of the main unit.
This section only applies to VUN-200 Temperature Probe. Please refer to section
4.2 for the cleaning of Heater Connection Cable and VUN-300 Connection Cable.
a. Rinse the temperature probe in clean water and a mild non-abrasive detergent, and
remove any dirt with a soft brush.
b. Flush the temperature probe with water.
Warning: Do not immerse the probe connector, which connects to the main
unit, in any liquid when washing and soaking the temperature probe.
c. Dry the temperature probe with a clean, dry cloth.
d. Soak for 3 minutes with 75% ethanol.
e. Dry the temperature probe with a clean lint free cloth.
f. Pack the temperature probe in clean plastic bag.
* The 75% ethanol disinfection solution can be replaced by following disinfectants:
1. Health Essence Disinfecting Effervescent Tablets
2. CIDEX OPA
* Follow the instructions supplied by the manufacturer of the disinfectant.
*