17
18
3) Connect Heated Breathing Circuit to the chamber.
4) Connect the heater connection cable and Temperature probe as indicated.
• Blue Port – Temperature Probe
• Red Port – Heater connection cable
NOTE:
• Do not allow the heated breathing circuit to come into direct contact with the
patient’s skin for prolonged periods of time.
• Do not use any accessory which is not recommended by Vincent Medical in this
manual.
If an integrated heated circuit with integrated temperature probe is used, VUN-300
Connection Cable shall be used in place of Heater Connection Cable and VUN-200
Temperature Probe. The connection is shown as below:
5) Select patient interface
The following table shows the dew point temperature and flow settings associated
with the relevant patient interfaces. Select the correct patient interfaces as per the
guidance of physician or healthcare professional.
Patient Interface
Nasal Cannula
(Infant)
VINC-01 (S)
VINC-02 (M)
VINC-03 (L)
VINC-04 (XL)
VANC-01 (S)
VANC-02 (M)
VANC-03 (L)
31~34
31~34
31~34
31~34
2~8
2~8
2~10
2~10
31~37
31~37
31~37
10~50
10~60
10~60
Nasal Cannula
(Adult)
Model
Dew Point Temperature (°C) Flow Rate (L/min)
* In low ambient temperature conditions, the dew point temperature may not reach 37°C
when the flow is high. User is advised to reduce the target flow setting to achieve the desired
dew point temperature.
Warning: The patient interfaces are all single use products. Do not reuse or use in
multiple patients as this may lead to cross-contamination. The patient interfaces must
be changed frequently to avoid infection.
Warning:
•
Replace chamber if water fill exceeds maximun level.
•
Disconnect water supply when machine is not running.
•
Do not use tap water, saline water or glucose water. Use sterile water for inhalation only.
•
Do not fill water through chamber outlet.
•
Do not tilt machine.
•
Do not use chamber for more than 14 days.
•
The chamber are all single use products. Do not reuse or use in mutiple patients.
SD card slot
Oxygen inlet port
Oxygen tube
Flowmeter
Heater
connection
cable port (red)
Heater
connection
cable
Temperature
probe port (blue)
Temperature probe
Autofeed humidification chamber
STERILE
WATER
Water Bag
STERILE
WATER